Breathing Pressure Monitoring
13-14
Part Number: 4116841-001
Narkomed MRI-2 Operator’s Manual
Rev: E
APNEA-P
ALM STBY
(LO)
Any time that the apnea pressure alarm (threshold pressure alarm limit)
has been set to Standby, the LO message
APNEA-P ALM STBY
appears in
the Alarm window.
BR PRS ERR
(LO)
If the Narkomed MRI-2 detects an internal electronic failure that would
prevent proper operation, the LO message
BR PRS ERR
appears in the
Alarm window. If this happens, contact an authorized representative of
DrägerService.
CONTNG
PRES (HI)
If the measured breathing pressure remains above the threshold pressure
alarm limit for more than 15 seconds, the breathing pressure display area
is cleared, the HI message
CONTNG PRES
appears in the Alarm window,
and an intermittent audible alarm sounds.
When the measured breathing pressure drops below the threshold
pressure alarm limit, alarm annunciation ceases.
O2 MON ERR
(LO)
If the Narkomed MRI-2 detects an internal electronic failure that would
prevent proper operation, the LO message
O2 MON ERR
appears in the
Alarm window. If this happens, contact an authorized representative of
DrägerService.
PEEP > 4
(LO)
Any time the monitor measures a PEEP of 5 cm H
2
O or greater, the LO
message
PEEP > 4
appears in the Alarm window.
When the measured PEEP drops below 5 cm H
2
O, the LO message is
cleared from the display.
PEEP > 25
(ME)
Any time that the monitor measures a PEEP of 26 cm H
2
O or greater, the
ME message
PEEP > 25
appears in the Alarm window and an intermittent
audible alarm sounds.
Alarm annunciation ceases when the measured PEEP drops below 26 cm
H
2
O. Also, an
APNEA
or
CONTNG PRES
alarm condition will clear this
alarm condition.
SUB ATM
PRES (HI)
If the measured breathing pressure falls below -10 cm H
2
O, the HI
message
SUB ATM PRES
appears in the Alarm window and a continuously
repeating audible alarm sounds.
This alarm condition is cleared when the sensed pressure rises above -10
cm H
2
O. However, the alarm message is extended for 5 seconds to allow
the recognition of a momentary subatmospheric pressure condition.
Summary of Contents for Narkomed MRI-2
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 6: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 66: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 74: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 80: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 84: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 88: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 116: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 140: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 172: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 176: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 177: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...