7. Temperature measurement
All settings concerning temperature measurement and
compensation are found in the submenu IN/OUT
=>TEMP. Select the symbol IN/OUT in the main menu,
and TEMP in the submenu.
Type
Select Pt100 if you have connected a Pt100, or Pt1000
if you have connected a Pt1000. Other temperature
sensors cannot be used.
Correction
If the sensor is connected with a 2-wire cable, slight
deviations might occur between measured and real
temperature. Check the temperature once during
installation,
and
calibrate
the
temperature
measurement if necessary by entering a correction
value between -10°C and +10°C.
Note
If the mode is manual the input for the correction value is blocked.
Reference
The reference temperature is the temperature to which the measurement is
compensated.
Mode
In the automatic mode, the measured temperature values are used to compensate
the temperature influence on the measurement, and in the manual mode, a manual
temperature setting is used.