18.8. Events with Popup messages
In some events, a window pops up, displaying a message. Just note its contents, or
follow the instruction. To close the window, you have to acknowledge with OK.
Popup message
Cause
Suggested remedies
SD card error
The instrument cannot
use the SD card
No SD card
Invalid format
Invalid file
SD card full
Check if an industry-standard SD card has been plugged into the slot, and that it was
plugged properly. If you want to load settings or update files, check that the files are
available on the SD card directly and not in a subfolder. Delete and restore the files,
if necessary. If you want to store settings on the SD card, check if there is enough
storage space left on the card.
This message appears also if you remove the card without deactivating the logger
first.
Popup message
Cause
Suggested remedies
No appli
The software files cannot
be loaded
Please renew the 2 data
files on the SD card and try
again.
For updates, both files have to be stored on the SD card, not in a subfolder. Check
that the files are available, and delete and restore them if necessary.
Popup message
Cause
Suggested remedies
Memory
stop
data
logging
The SD card is fulll.
SD card full, data logging
stopped.
If you have set the logger to STOP mode, then the instrument stops the logging when
the SD card is full.