5. How to use device
5.1 Device description
5.2 Start-up device
For operation take out the instrument from the packaging, remove the
display foil. Before switching on the instrument connect the probe/s to
the instrument and check the batteries (2 x 1,5V AA size) whether they
are inserted. The devices type 4005 and 4015 have two probe sockets
(channel 1 on the right bottom and channel 2 on the left bottom).
By operating the ON/OFF-button the instrument switched on or off. After
switching on the instrument shows for one second all display segments,
then it starts operating in measurement mode.
The instrument has a built-in Mini-USB-Port. Through our optional
DE-Graph Windows Software (Order Code 5090-0081) it is possible to
transfer the measurement to a PC (online documentation). By a USB
connection the instrument will be power supplied, too. This product is
exclusively intended for the field of application described above.
It should only be used as described within these instructions.
Unauthorized repairs, modifications or changes to the product are
prohibited and void any warranty!
Fig. A
(see page 2)
1. LCD-Display see Fig. D
2. USB-connector see Fig. B
3. ON/OFF-button
4. CLEAR-button
5. Ports
5. a Channel 1 (Pt100)
5. b Channel 2 (Pt100)
5. c Channel l 1 (type K)
5. d Channel 2 (type K)
6. MEMOUT-button
7. MEMIN-button
8. MODE-button
9. Battery case see Fig. C
Fig. D
(see page 3)
1. Measurement result
line 1 / channel 1
2. Measurement result
line 2 / channel 2
3. Display differential
temperature (T1-T2)
4. Status line
HOLD = keep the last
measurement result
on the display
5.3 Factory presets
• Temperature unit in degree Celsius °C
• Power-Save Mode: on (The instrument switches off automatically
after appr. 15 minutes)
5.4 Functions
5.4.1 Measuring unit switching °C Celsius and °F Fahrenheit
To change the measuring unit hold the ON/OFF-button and
MEMOUT-button together during switching on procedure until a
measurement appears.
5.4.2 Calibration
The instrument offers an easy calibration function to adjust the
temperature at one point – to compensate sensor tolerances.
Push once the MODE-button. On the bottom of the display you will
see HOLD and the displayed measurement value is “frozen”.
By using now the MEMIN-button the instrument switch to
CAL-mode. On the instrument appears 00.0
The first 0 is blinking. By using the MEMIN-button you increase the
blinking figure and by using the MEMOUT-button you decrease the
measurement value. By using the MODE-button you will switch to
the next figure or you will end the adjustment.
Note:
The mentioned offset will add or subtract (negative value)
from the measurement.
AVG = average value,
MAX = maximum value
MIN = minimum value,
5. Battery indication
6. Adjustment function active
7. Unit