2. Connect the console to the

12 VDC power supply

and turn the console

ON

.

THIS MUST BE DONE BEFORE

CONNECTING THE CONSOLE TO THE COMPUTER.

3. Connect the

Fi-Wi Headstage Charger

to the

USB Docking Port

using the USB-A (male/female) cable. Install the

headstage(s) onto the charging port.

4. Connect the console to the computer using the

USB-A/USB-B Cable

.

5. When the console is

OFF

, disconnect the

USB-A/USB-B Cable

before turning it back

ON

. Once the console is

ON

,

reconnect the cable.

2.1.2 Software setup

The following section details how to set up a measurement sequence using

Doric Neuroscience Studio

. Further detail for

each software feature is found in section

3

.

Figure 2.3:

Antenna Channel Configuration

1. With the console properly connected, open the

Doric Neuroscience Studio

. Once open, the electrophysiology con-

sole tab (Fig.

2.3

will appear. Select the

Configuration

tab in the

Control & Settings

box, then select the

Add

configuration

button to open the

Channel configuration

window.

2. In the

Channel configuration

window (Fig.

2.4

), there are 2

Antenna

channels available. When one is selected,

the channel can be configured.

• The

Acquisition settings

box (Fig.

2.4a

-1) is used to define how electrophysiological data is acquired. This

includes triggering options, as well as frequency filters.

• The

LED Options

box (Fig.

2.4a

-2) is used to define the cannula LED current and the current baseline.

• The

Headstage settings

box (Fig.

2.4a

-3) is used to define the pulse sequence emitted by the cannula LED.

While the standard pulse sequence is square, the

Smoothing Edges

function can allow specific rise/fall times.

3. Once your channel is configured, it can be controlled using the

Acquisition

control tab and the tools in the

Acqui-

sition view

.

• Once the headstage is charged and off the charging station, click the

Pair/Unpair

button (Fig.

2.4b

-1), which

will connect it to the console. Its

Unique ID

will be indicated right above the button. The wifi

Channel

is

shown beside the

Unpair

button; if multiple headstages are in use, each much have a different channel sepa-

rated by at least 10 increments.

• Once the headstage is paired, the LED sequence is triggered by the beginning of an acquisition sequence.

The sequence progression can be seen on the

LED Trigger

window (Fig.

2.4b

-2).

• Once everything is ready, the

Live/Record

buttons (Fig.

2.4b

-3) will start the acquisition sequence.

• With the acquisition sequence started, recorded signal will be shown in the

Graphs Box

(Fig.

2.4b

-4).

Chapter 2. Operations Guide

9

Summary of Contents for Fiberless and Wireless Optogenetically Synchronized Electrophysiology System

Page 1: ...Fiberless and Wireless Optogenetically Synchronized Electrophysiology System User Manual Version 2 1 0...

Page 2: ...Guide 8 2 1 System Setup 8 3 Doric Neuroscience Studio 12 3 1 Fiberless and Wireless Optogenetically Synchronized Electrophysiology 12 3 2 Acquisition Console 17 3 3 Electrophysiology Analysis Module...

Page 3: ...ure 1 1 Optogenetically Synchronized Electrophysiology System Layout The system uses a Fiberless Wireless Headstage to connect the Electrophysiology Console and the Fi Wi Opto electric Can nula using...

Page 4: ...electrophysiological signal This is done using the Intan RHD 2132 chip Its secondary use is as an LED driver It sends pulse sequences to the stim ulation LED contained in the cannula The headstage als...

Page 5: ...nsole Figure 1 4 Electrophysiology Console Front The Electrophysiology Console Fig 1 4 allows control of the different elements as well as reception and transmission of information from the control co...

Page 6: ...ing Port its wireless connection is shut down The Status Lights on the side of the charger show the status of connected headstages The one closest to the connector is blue when the charger is ready fo...

Page 7: ...ing elements The Stereotaxic Clamp is used to attach the holder to a rod in a stereotaxic apparatus TheHeadstageConnectorisusedtoholdaFi WiHeadstage Aheadstageinthispositionwillbeabletosend receive si...

Page 8: ...n of interest of the animal s brain Fig 2 1 The process of implantation is not described in this document Figure 2 1 Implanted Fi Wi Opto electric Cannula 2 1 1 Electrophysiology Console Setup Figure...

Page 9: ...ettings box Fig 2 4a 1 is used to define how electrophysiological data is acquired This includes triggering options as well as frequency filters The LED Options box Fig 2 4a 2 is used to define the ca...

Page 10: ...headstage onto the USB Charger so the antenna is on the position indicator The console must be ON for the headstage to recharge which requires 60 min The headstage cannot connect wirelessly while bein...

Page 11: ...bed in section 1 1 2 7 After use place the Fiberless and Wireless Headstage back onto the USB Charger to reset the wireless connec tion 8 To ensure optimal transmission The experimental subjects shoul...

Page 12: ...ncorporates different functions to control our Doric Fiberless Wireless OSE system The core of the system is the Fiberless Wireless Headstage This device controls a Wireless cannula containing several...

Page 13: ...ecting the channel icon will display the parameters of each respective Channel type 2 The Wireless settings Fig 3 2 are used to select recording and transmission settings a The Antenna selection Fig 3...

Page 14: ...put one pulse every 2000 ms period iii The Time ON Duty Cycle Fig 3 2 sets the time in ms or the duty cycle in for each pulse The Time ON must be lower than 1 frequency 0 005 ms while the Duty cycle m...

Page 15: ...ed on the window occupying a rectangular box Each Channel box shows a number of basic elements described in section 3 2 3 Elements uniquely associated with the Antenna channel are described here Figur...

Page 16: ...shows the channel currently in use by the headstage When using two head stages the channel must be different for each headstage f The Pair button is used to pair an active headstage to a console When...

Page 17: ...It enables control over the excitation light pulses or the sinusoidal waveform trig of an ex ternal source i e Doric LED driver with 4 TTL and 4 analog voltage outputs The module interface displays r...

Page 18: ...type is currently being modified The channel can be changed by selecting a new one from the drop down list b The Mode identifies the type of signal sent for output channels or the way the signal is m...

Page 19: ...ces Each numbered channel corresponds to the same number digital channel on the console Pulse se quences have different parameters depending on the Channel Mode which can be Continuous or Square for o...

Page 20: ...can also be converted to a Duty Cycle which indicates the of the period the pulse duration corresponds to 3 The Pulse s per sequence set the number of pulses per sequence If it is set to 0 the number...

Page 21: ...ol lowing elements appear in the Sequence Options box 1 The Starting Delay Fig 3 10 defines the time between the activation of the pulse sequence and the begin ning of the signal 2 The Time ON Fig 3 1...

Page 22: ...conds period 2 The Time ON Fig 3 11 defines the length of a single pulse This time can also be converted to a Duty Cycle which indicates the of the period the pulse duration corresponds to 3 The Maxim...

Page 23: ...tes a sinusoidal pulse sequence with peaks at V and V This includes all sequence options as the CW and Square mode save the following The Time ON option is changed for the Phase option Fig 3 12 This a...

Page 24: ...ion of each Mode is described here The Linear channel mode Fig 3 13 allows the direct measurement of signal received by a channel The linear mode has the option of a frequency filter found in the Filt...

Page 25: ...a Filter Tab b Response Tab Figure 3 14 Linear Mode Filter Options 7 The Response Tab displays the filter response attenuation according to frequency Chapter 3 Doric Neuroscience Studio 25...

Page 26: ...hannels while periodically saving recorded data to the disk This mode is recommended for long measurement sequences Figure 3 16 Time Series Acquisition Timing Diagram 3 The Time Series button Fig 3 15...

Page 27: ...nt indexation number used for multiple files saved during the same measurement session All this information is summarized in the Target File box The Decimation Settings allow the reduction of file siz...

Page 28: ...ch current channel 4 The Load configuration button Fig 3 19 allows a console configuration in doric format to be loaded View Tab The View Tab Fig 3 20 is used to modify the presentation of graphs in t...

Page 29: ...Mode Trigger Mode and Trigger Source if applicable Selecting the button while a window is open will close the window Figure 3 22 Legend Window 4 The Show Notes button Fig 3 20 will open the Note View...

Page 30: ...ponding to that on the console This name can be modified in the Graph options window 2 The Edit button Fig 3 24 allows the editing of channel parameters opening the channel configuration windows For f...

Page 31: ...than the default none may result in slower display and perfor mance degradation 4 The Graph Fig 3 24 box displays the graph of a channel with Voltage or State as the vertical axis and Time as the hor...

Page 32: ...b allows the loading and saving of data Figure 3 28 File Tab The Load data button allows the loading of a csv file The file must contain time and signal data The Load notes allows the loading of a jso...

Page 33: ...in Set zoom The Toggle graphs button opens the Show hide graphs window Fig 3 31 Any checked data sets will be dis played in the Graph box Figure 3 31 Show hide Graphs Window The Toggle cursor button w...

Page 34: ...e data set to be filtered the filter window Fig 3 34 is opened This allows a choice of filter type and variables Figure 3 34 Filter Window The Filter specification allows the selection of a filter mod...

Page 35: ...The Window before after threshold defines the time recorded in ms before after the point identified as a spike The Minimum time between events defines the minimum time in ms between two different spi...

Page 36: ...ter plot of all spikes over time with each color corresponding to a single cluster This window can be kept active over multiple uses of the Spike finder with all raster plots displayed within Figure 3...

Page 37: ...t spikes associated with each individual unit Figure 3 41 Unit box 5 The Timeline box shows the original graph with each spike identified Each color indicates a given unit as defined in the legend of...

Page 38: ...r core diameter 200 or 400 m Table 4 2 Fiberless Wireless Headstage General Specifications SPECIFICATIONS VALUE NOTES Dimensions 9 6 x 14 9 x 18 6 mm Mass 2 8 g Including 40 mAh battery Transmission F...

Page 39: ...00 Hz 26 1 V 18 5 bits 27 0 V bit 6 0 kSps C 6 041 Hz 166 s 3 100 Hz 46 0 V 17 7 bits 47 0 V bit 12 kSps C 12 166 Hz 82 s 6 300 Hz 120 V 16 3 bits 123 9 V bit Table 4 5 Console Recommended Environment...

Page 40: ...urn instructions This warranty will not be applicable if the unit is damaged or needs to be repaired as a result of improper use or operation outside the conditions stated in this manual For more info...

Reviews: