Section 2. Battery Operation 

 

Battery Installation 

The indicator can be powered by a set of six AA alkaline batteries.  These 
batteries can perform over 25,000 weighments of typical use with the standard 
Automatic Shutoff Timer settings.   
 
The battery compartment is located on the backside of the indicator. Press the 
locking tab to release battery compartment cover. Completely remove all the 
batteries from the indicator and install the new set of batteries.  Be sure the 
batteries are oriented properly.  Reinstall the battery compartment cover to 
complete the battery installation. 
 
 

Battery Indicator 

The battery indicator displays the remaining battery life available.  A low battery 
status is also indicated by two beeps when a weight is held (If Parameter 

BH

 is set 

to 

on

.)  This indicator will not be displayed if the scale is powered by its wall 

transformer. 
 
 
 
 

 

 

 
 

Summary of Contents for DS6150

Page 1: ...Model DS6150 Remote Indicator Scale Technical Manual Doran Scales Inc 1315 Paramount Pkwy Batavia IL 60510 USA 1 800 264 4107 www doranmedical com MAN0315 Revision 0 1 2 12 2014...

Page 2: ......

Page 3: ...5 Parameter Setup 12 Entering and Exiting Setup Mode 12 Calibration Mode Entry 12 Reset All Parameters to Default Settings 13 Automatic Shutoff Timer 13 Average Aperture 14 Data Output 14 EMR Format...

Page 4: ...y not be covered by the warranty Please observe the following precautions to insure years of trouble free service from your new scale Do not use sharp objects to press any of the buttons Do not pick u...

Page 5: ...2 Wall Mounting If mounting the indicator to a wall refer to the following diagram Place two screws 4 09 apart 65 above the floor The screw heads should be approximately 0 33 in diameter...

Page 6: ...to release battery compartment cover Completely remove all the batteries from the indicator and install the new set of batteries Be sure the batteries are oriented properly Reinstall the battery comp...

Page 7: ...4...

Page 8: ...r 3 seconds to turn off the indicator The indicator has an adjustable Automatic Shutoff Timer that will turn off the indicator when not in use Zero When the ZERO button is pressed the scale will reset...

Page 9: ...e Hold Indicates a stable weight has been achieved and held on the display This indicates an accurate weight has been achieved Note Weights below 4 5 pounds will not be held Average Hold Indicates tha...

Page 10: ...mber and Revision display Pressing and holding the REWEIGH button for 3 seconds will perform a display test The software number Su179 and the revision r 3 2 or higher are displayed first After display...

Page 11: ...econds BMI Operation 1 With the held weight displayed press the BMI button to enter the height value Height entry starts at 5 6 0 or 168 0 representing 5 foot 6 inches or 168 centimeters NOTE Height u...

Page 12: ...or by using your own body weight Two Point Calibration Procedure 1 When the setup mode is entered the display reads CAL n 2 Press the button to change the display to CAL y 3 Press BMI 4 The display r...

Page 13: ...e sure vibration and air currents are not present 9 Press BMI to perform a zero point calibration The display will count down to zero The counter will reset if too much motion Er nn is detected 10 The...

Page 14: ...alibration has detected a negative range If this error persists consult the raw counts table The following table shows the acceptable Raw Counts for no load and full load Raw Counts can be viewed thro...

Page 15: ...gation 1 Press BMI to scroll forward to the next parameter 2 Press or buttons to change the current parameter setting 3 Press Reweigh to save any changes and exit setup mode Calibration Mode Entry Cal...

Page 16: ...ration is performed Reset All Parameters to Default Settings rSt Default parameters n No Parameters are not defaulted y Yes Set all parameters to the factory default values when BMI is pressed Calibra...

Page 17: ...n a request command is received through the RS 232 or USB port A 5 Transmits an EMR string after the STABLE HOLD or AVERAGE HOLD weight is displayed default A b Transmits an EMR string after the Heigh...

Page 18: ...disabled default LB OZ Units Enable Disable lb oz units Enable Disable on lb oz weighing units active of lb oz weighing units disabled default Inches Units Enable Disable Inches Height units Enable D...

Page 19: ...disabled y Beep enabled default Battery Voltage Level DC Displays current battery voltage DC Battery Bar Indications Empty Battery Symbol 5 1 5 9 V One Bar 6 0 6 9 V Two Bars 7 0 7 9 V Three Bars 8 0...

Page 20: ...ation of that system These are Baud Rate Data Bits and Parity The baud rate determines how fast the data is sent from the scale The sending and receiving units must be set to the same baud settings Re...

Page 21: ...ng the Reweigh button Welch Allyn Spot protocol VU 2400 7 EVEN 1 Command RXD Computer to Scale Scale Output Responds TXD Scale to Computer Description DR cr lf mot sp sp weight uu cr lf Current Weight...

Page 22: ...will wait until stable then zero cr or W bmi height pol gross uu sp G BMI uu sp T pol net uu sp N cr lf Current Display Weight Active Stable Hold or Average Hold mot G or N stable g or n motion U Scal...

Page 23: ...ble character cr ASCII 0dh Carriage return non printable character lf ASCII 0ah Linefeed non printable character sp ASCII 20h space mot GR for stable gr for motion or SD for stable S_ for motion pol m...

Page 24: ...6 5 W x 2 3 H Power 12VDC AC Adapter Negative Center Optional 6 AA Alkaline Batteries Displayed Units lb kg Capacity 500 x 0 2 lb or 230 x 0 1 kg Construction Painted Mild Steel Frame Stainless Steel...

Page 25: ...zing Motion annunciator is off After pressing the ZERO button the scale should zero as soon as it becomes stable Weight readings are not correct Verify the scale calibration with an accurate set of te...

Reviews: