![Doran Scales DS6150 Technical Manual Download Page 13](http://html1.mh-extra.com/html/doran-scales/ds6150/ds6150_technical-manual_2521652013.webp)
10
14. Press BMI to accept the span weight value. The display will count down to zero
and return to the weigh mode.
Multiple Point Calibration Procedure
1.
When the setup mode is entered the display reads
CAL n
.
2. Press the
▲arrow button to change the display to
CAL y
.
3. Press BMI.
4. The display reads
CAP
,
then the current capacity value in lb or kg.
5. Press the
▲or ▼arrow to select a capacity from a table, example 500 lb (230
kg). Press the UNITS button to toggle between lb or kg capacity units.
6. Press BMI to save current capacity and advance to the zero point calibration.
7. The display reads
CAL 0
.
8. Remove all weight from the scale platform.
NOTE: During calibration be sure vibration and air currents are not present.
9. Press BMI to perform a zero point calibration. The display will count down to
zero. The counter will reset if too much motion (
Er nn
) is detected.
10. The scale will display
500.0
. Representing a 500 lb span point calibration.
11. Press Zero button, to enter multipoint calibration
12. The scale will display
100
lb (
45
kg)
13. Place 100 lb (45 kg) onto the scale.
14. Press BMI. The display will count down to zero.
15. The scale will display
200
lb (
90
kg)
16. Place 200 lb (90 kg) onto the scale.
17. Press BMI. The display will count down to zero.
18. The scale will display
300
lb (
135
kg)
19. Place 300 lb (135 kg) onto the scale.
20. Press BMI. The display will count down to zero.
21. The scale will display
400
lb (
180
kg)
22. Place 400 lb (180 kg) onto the scale.
23. Press BMI.
24. The display will count down to zero and return to the weigh mode.