12
Section 5. Parameter Setup
The indicator has many adjustable parameters which control the scale’s operation. The
parameter settings shown in bold are the default settings.
Entering and Exiting Setup Mode
To enter setup mode from the front panel, press and hold REWEIGH and ZERO
for 5 seconds. Release both buttons when the display reads
CAL n
.
To exit the setup mode and save changes, press REWEIGH or BMI to scroll to
the end of parameter menu.
Note: If no buttons are pressed within 30 seconds,
the scale will automatically exit the setup mode.
Setup Mode Navigation
1. Press BMI to scroll forward to the next parameter.
2. Press
▲ or ▼ buttons to change the current parameter setting.
3. Press Reweigh to save any changes and exit setup mode.
Calibration Mode Entry
Cal
Enter Calibration Mode
(60 second timer)
N
Do not enter calibration
y
Enter Calibration
(see Scale Calibration section)
Cap
Capacity
(60 second timer)
Use the UNITS button to toggle
between lb or kg capacities.
45
(lb)
20
(kg)
45 x 0.02 lb or 20 x 0.01 kg
50
(lb)
25
(kg)
50 x 0.02 lb or 25 x 0.01 kg
100
(lb)
45
(kg)
100 x 0.05 lb or 45 x 0.02 kg
500
(lb)
230
(kg)
500 x 0.2 lb or 230 x 0.1 kg
550
(lb)
250
(kg)
550 x 0.2 lb or 250 x 0.1 kg
600
(lb)
270
(kg)
600 x 0.2 lb or 270 x 0.1 kg
800
(lb)
360
(kg)
800 x 0.2 lb or 360 x 0.1 kg (default)
1000
(lb)
450
(kg)
1000 x 0.5 lb or 450 x 0.2 kg