Instruction Manual 

※ Read this instruction manual before using the product in order to achieve maximum 
performance. 
※ Keep this instruction manual within your reach after reading so that it can be used at any time. 
※ Warranty is on the last page of this manual. Please read the contents and keep it carefully. 

HAND-HELD ELECTROSTATIC VOLTMETER 

ARS-H002ZA

 

28, Namyang-ro 930 beon-gil, Hwaseong-si, Gyeonggi-do, Korea 
Tel 82.31.299.5453 Fax 82.31.357.2610 

Summary of Contents for ARS-H002ZA

Page 1: ...his instruction manual within your reach after reading so that it can be used at any time Warranty is on the last page of this manual Please read the contents and keep it carefully HAND HELD ELECTROSTATIC VOLTMETER ARS H002ZA 28 Namyang ro 930 beon gil Hwaseong si Gyeonggi do Korea Tel 82 31 299 5453 Fax 82 31 357 2610 ...

Page 2: ... Page 9 Attention 16 Page 10 Trouble shooting 17 Page 11 Warranty 18 Page 2 DONG IL TECHNOLOGY LTD CONTENTS 7 1 Power supply Battery use 7 2 Measuring set up 7 3 Setting zero point adjustment 7 4 Electrostatic mode 7 5 Ion Balance mode 7 6 HOLD function 7 7 Automatic Power OFF function 8 1 Installation and connection of monitoring program 8 2 Monitoring program method 8 3 Case of disconnecting to ...

Page 3: ...e is designed with rotary integrated structure so it is more convenient for use and storage of the product Real time Monitoring System by PC It is able to check the measured data in real time and save the data via USB port Monitoring program is provided free of charge DONG IL TECHNOLOGY LTD Welcome to become a customer of DIT Static electricity can cause lower production efficiency such as damage ...

Page 4: ...onse Time 1sec Function Automatic power OFF Alarm for power OFF Zero setting Electrostatic Ion Balance mode HOLD function Display HTN LCD Digit and Gradation display Material Conductive ABS Operating time Max approx 30hours Communication USB Communication Ambient Temperature 10 40 50 104 Ambient Humidity 0 60 RH Non Condensing 4 2 Specification DONG IL TECHNOLOGY LTD Specification can be changed w...

Page 5: ...his instruction manual before using the product EARTH CABLE ARS H002ZA main body Battery USB CABLE Instruction Manual Download instruction manual on web site 6LF22 or 6LR61 DONG IL TECHNOLOGY LTD Battery is provided in main body This image could not be matched for real battery 1 Battery socket is disconnected to prevent battery consumption during distribution Please open the rear of the product an...

Page 6: ...r measuring ION BALANCE LCD PANEL Display measured value Menu status Power status etc POWER BUTTON Turn power ON OFF ZERO BUTTON Set up to zero adjustment HOLD BUTTON Hold the measured value IB BUTTON Choose the Electrostatic mode or Ion Balance mode LABEL 1 Caution LABEL 2 Information of product and company BATTERY BOX COVER Cover of the battery slot ...

Page 7: ...real time monitoring with PC Ground Terminal Earth Cable terminal Plate rotary button Both buttons should be pulled down when rotating the plate for measuring ion balance LED for focusing Two LED guide lights help position the voltmeter at 25mm distance from a charged test object Sensing Hole Sensor for static electricity Main body left Main body upper ...

Page 8: ...8 Front View Top View Side View 5 Dimension DONG IL TECHNOLOGY LTD ...

Page 9: ...9 6 Performance DONG IL TECHNOLOGY LTD Display data according to charged voltage Display data according to charged voltage Precision mode Expansion mode Charged Voltage kV Charged Voltage kV ...

Page 10: ...rthed objects with an ground cable 2 It is recommended to wear Earth Ring or ESD gloves to avoid a static electricity from human body 3 Press POWER button to operate the product apart from a charged test objet or ionizer 4 Check the display is same as below after power on If Earth cable is not connected inexact values could be measured Please use the product after checking the product earthed Init...

Page 11: ...measures the voltage or electrostatic level on objects with an integral sensor on Electrostatic mode 1 Check to set the electrostatic mode as below 2 Place the device at right angle to test object as picture below Adjust the distance until two LED light marks become the one on the target Description Polarity of voltage Digit 2kV 2kV 20kV 20kV Bar graph approx 2kV 1 bar L R Unit of measure kV An er...

Page 12: ...ricity is over measurable range In this case please stop measuring because integral sensor can be damaged HHH over voltage measurable range 20kV LLL over voltage measurable range 20kV Description Ion balance mode Polarity of voltage Digit 300V 300V Bar graph approx 30V 1 bar L R Unit of measure V 7 Operating Instruction DONG IL TECHNOLOGY LTD Warning 1 Make sure to pull down the plate rotary butto...

Page 13: ...mode Ion balance mode When the value on DISPLAY is hard to be checked this function holds the value on DISPLAY So it is convenient to check the value even after changing the position 1 Follow the operating instruction for measurement 2 Press HOLD button on main body once when to measure the value 3 When positioning the product with measurable distance the value is fixed and HOLD is displayed on LC...

Page 14: ...r 3 seconds Then Automatic Power OFF function is turned off and it is displayed on LCD as below 3 If not to use this function the product keep continuously power on until user turns off power Please note the battery charge status 4 If user want to reactivate this function press HOLD button over 3 seconds Then blinking of the battery charge status stops and Automatic Power OFF function is reactivat...

Page 15: ...on our web site 2 Install the USB driver FT232RL to PC 3 Connect ARS H002ZA and PC with a provided USB cable 4 Check the port number of driver available on Microsoft Management Console 1 Check the status of USB cable connecting ARS H002ZA and PC 2 Check that the port number of USB driver is same as the port number of software 3 Check to set 9600 baud on monitoring program settings 4 Check to set 9...

Page 16: ...st 10 Do not touch or put sharp objects to the sensing hole 11 When measuring ion balance do not face the sensing hole to the object without the sensing plate 12 Do not press buttons on main body with excessive force 13 Keep the sensing plate at a humidity under 60 RH because of its high insulation recommended to use desiccator or plastic bag and desiccant 14 The sensing plate should maintain its ...

Page 17: ...o manufacturer or sales agents Unusual value on DISPLAY 1 Check that ground cable is connected properly 2 Check the settings for measurement mode Error is bigger than real electrostatic value 1 Check the measurement distance 25mm 0 5mm 2 Check the any other object affecting its performance 3 If there is no problems above contact to manufacturer or sales agents LED light is OFF 1 Contact to manufac...

Page 18: ...a recommendation and users are responsible for understanding the product s specification and judgment of suitability of usage 3 Direct or indirect damage led by the product malfunction DONG IL TECHNOLOGY LTD 28 Namyang ro 930beon gil Hwaseong si Gyeonggi do Korea Tel 031 299 5453 Fax 031 357 2610 Technology Market Solution DIT Technology Application Experts are focused on delivering solutions for ...