10. Trouble shooting
17
※ If you cannot solve the problem with the above solution or if you have other problems
not described above, please call manufacturer or leave a message on our website.
☞ DIT Quality Assurance Team TEL) +82-31-299-5464
Problems
Check points
Power is not supplied
1) Check the battery connection
2) Check the battery charge status
3) If there is no problems above, contact to
manufacturer or sales agents.
Unusual value on DISPLAY
1) Check that ground cable is connected properly
2) Check the settings for measurement mode
Error is bigger than real electrostatic
value
1) Check the measurement distance [25mm(±0.5mm)]
2) Check the any other object affecting its performance.
3) If there is no problems above, contact to
manufacturer or sales agents.
LED light is OFF
1) Contact to manufacturer or sales agents
Burning smell during operation
1) Remove the battery and contact to manufacturer or
sales agents
ETC.
1) Contact to manufacturer or sales agents
DONG IL TECHNOLOGY LTD