Technical User Guide

DOL 51

www.dol-sensors.com

3

140299 DOL 53 – test gas adapter

Used for gas test of DOL 51 and DOL 53.

3 Installation

Mount the sensor in the mounting lug at the top of the sensor, filter facing down. The offset from vertical must
be max. 15

o

 (see figure 1). DOL 51 must be mounted so that it is accessible in connection with maintenance

and must not be subjected to splash water.

Ensure free air passage around the sensor filter. Do not expose it to direct sunlight as this will affect the mea-
surements.

DOL 51 must warm up for half an hour before being ready to measure (see the section of LED indication).

The sensor must be placed in middle of the livestock house or in the center between walls/filters when used in
an air cleaner. Make sure there is an adequate distance to walls and air inlets.

DOL 51 must not be suspended from the electric cable.

Figure 1: DOL 51 must be mounted vertically with the sensor pointing down; +/- 15° at a location with low vibra-
tion and, as far as possible, stable temperatures.

Summary of Contents for DOL 51

Page 1: ...DOL 51 Ammonia sensor For other language variants of this document we refer to www dol sensors com or your local dealer dol sensors a s Agro Food Park 15 DK 8200 Aarhus N T 45 7217 5888 F 45 7217 5959...

Page 2: ...lement and filter can be replaced as required and the sensor element is supplied in a can for additional protection The sensor is meant for stationary mounting for continuous measuring of the ammonia...

Page 3: ...r passage around the sensor filter Do not expose it to direct sunlight as this will affect the mea surements DOL 51 must warm up for half an hour before being ready to measure see the section of LED i...

Page 4: ...shooting of all electrical equipment must be carried out by quali fied personnel in compliance with the applicable national and international standard EN 60204 1 and any other EU standards that are ap...

Page 5: ...e g 230 V AC applied electronics damaged Replace the DOL 51 housing DOL 51 housing is defective Replace the DOL 51 housing contact the sup plier Alternating GREEN and YELLOW Sensor is warming up to b...

Page 6: ...1 housing is defective Replace the DOL 51 housing Incorrect measurement on the control unit The dust filter is clogged Replace the dust filter Check the sensitivity with test gas Unexpected measuremen...

Page 7: ...th lukewarm water Check for mechanical damage Transport The sensor must be transported in the original packing in air traffic Otherwise the sensor may lose its measuring sensitivity 9 Technical data D...

Page 8: ...ensor can Warranty sensor element 1 years Expected lifetime of sensor element Not beyond the date on the sensor element type plate 1 years IP classification IP65 Cable length 6 m Cable conductor size...

Reviews: