6.3 End-of-Line Particulate Filter Replacement
End of line (point of detection) particulate filters which protect the DOD64-FTIR from damage are required on
all points including points not being monitored. Table 7.1 details the type of filter required for each gas.
Filters must be replaced on a regular basis as shown in the table. Filter orientation is not critical in either
application.
IMPORTANT:
All points require filtration to prevent dust accumulation in tubing and internal damage to the
DOD64-FTIR. Dust that collects in the tubing or the internal system may cause sample loss and inaccurate
gas concentration readings.
Table 7.1
Mineral Acids Filter
Part Numbers:
Blue Housing: 60009
Filter Membrane: 60010
Hydrides & Phosgene Filter
Part Number:
780248
Mineral Acids Disposable Filter
Part Number
2-800-013
6.4 Flow Adjustment
Each channel should be adjusted whenever a particulate filter is installed.
6.5 USB and Solid-State Storage Drive Replacement
It is highly recommended to keep a USB flash drive or solid-state inserted in the unit at all times. A general
fault is issued anytime a USB or solid-state drive is not inserted or full.
To insert or replace a USB drive follow these steps:
1.
Exit Analysis
2.
Go to the USB Drive removal screen on the main menu under History.
3.
Touch the 'remove' button to halt writing to the disk.
4.
Follow the procedures from section 6.3 to open the control panel access door.
5.
Remove the USB drive from the system and replace with new drive
6.
Wait 5 seconds for the system to initialize.
7.
Close and latch the control panel access door.