FTIR Manual
5.2.1 Faults and Events
Figure 5.3
Pressing the Faults/Events tab at the top of the Analysis screen will display the Faults/Events screen (figure
5.3), allowing you to view a detailed history of the machine’s faults
and/or events. You may select any day on
the cal
endar to view a log of that day’s events.
You must press inside the fault list in order to activate the
Page Up, Up, Down, and Page Down Functions.
Green
Events indicate instances when analysis mode was activated
Blue
Event display non-critical events, such as output testing or a new background
Yellow
Events indicate a general fault that does not remove the machine from analysis. This type
of fault will trigger the general fault output.
Orange
Events indicate a critical fault that brought the machine out of analysis mode. This type of
fault will trigger the critical fault output.
Red
Events indicate a gas alarm level had been reached. This type of event will trigger either
gas alarm 1 or 2 outputs.
In the event of a critical fault, the machine will stop analysis and display this screen. Clearing the fault and
resuming analysis is possible by pressing the Fault/Alarm Reset button. Check appendix D for a list of faults.
This screen can also be accessed from the Setup > History menu.