10
Thermal block format (modification X)
384-well microplate
Tube type (modification M)
0.2-ml test tubes for PCR (individual, in strips, 8
pieces each, or in plates 12 x 8)
Tube type (modification X)
384-well microplate, 24x16 wells of 45 mcl
volume
Reaction volume (modification M)
10-100 mcl (10-50 mcl recommended)
Reaction volume (modification X)
5-30 mcl
Temperature range
0 C-100°C
Temperature step
0.1°С
Temperature accuracy
±0.2°С
Temperature uniformity
±0.15°С
Average heating ramp rate within temperature
range of 0…100°С (modification M)
Maximum heating ramp rate within temperature
range of 0…100°С (modification M)
3.3°C/s
3.5°C/s
Average cooling rate within temperature range of
100...55°С (modification M)
Maximum cooling rate within temperature range of
100...55°С (modification M)
2.1°C/s
2.5°C/s
Average heating ramp rate within temperature
range of 0…100°С (modification X)
Maximum heating ramp rate within temperature
range of 0…100°С (modification X)
2.1°C/s
2.5°C/s
Average cooling rate within temperature range of
100...55°С (modification X)
Maximum cooling rate within temperature range of
100...55°С (modification X)
1.0 °С /s
1.5 °С /s
“Hot lid” temperature
105°С ±1°С
Heating and cooling method
Peltier elements
Excitation source
LED
Detector
CCD camera