DMC Therapy Plus User Manual Download Page 6

6

Therapy Plus | 

User’s Manual

Features

Specifications

Manufactured and tested according to:

IEC 60601-1, IEC 60601-1-2, IEC 62034, IEC 60601-
2-22 and IEC 60685-1.

Software version

12

Use temperature

10 °C - 30 °C

 

SAFETY - IMPORTANT PRECAUTIONS

•  Please read the whole manual before using the equipment.
•  Attention - Using other controls, setting or running not specified in this manual may result in the 
exposure to hazardous radiation.
•  Attention - Laser smoke may present living tissue particles.
•  As the equipment will be used by qualified professional, an additional specific training for its using 
is not required. However, the reading of whole manual before its using is recommended.
•  If the equipment is not used for some time, the battery shall be removed.
•  The replacement of the battery by a personnel with inadequate training might result in a danger 
(such as excessive temperature, fire or explosion).
•  Laser lights may cause eye injuries. People who are on site should protect themselves against laser 
lights emission. Therapy Plus supplies three green glasses (one for operator, the other for assistant 
and the other one for patient). Only glasses supplied by DMC should be used with the equipment.
•  Never look directly to laser light. Never direct it for any person, except he/she is on treatment.
•  Bright surfaces may reflect laser light towards the eyes.
•  Never radiate tumor processes directly, the laser can stimulate them.
•  Never radiate an undiagnosed injury.
•  Only skilled professionals should handle the equipment. An inappropriate use may bring irrevers-
ible injuries.
•  Only components mentioned in this manual should be used together with equipment.
•  This equipment cannot be used with power cable not supplied by DMC because it may provoke an 
increase of emission or decrease of immunity of equipment.
•  To avoid risks of shocks, the equipment should be connected only in grounded outlets.
•  Do not connect power cable in difficult-to-access outlets.
•  Disconnect the power cable from outlet by plug.
•  No change should be made in the equipment.
•  Do not obstruct equipment air input/output.
•  Do not apply any protector film on handpiece in order to obstruct air input/output.
•  Do not install the equipment directly on sunlight, dusty places or mechanical vibration.
•  There are risks of fire and/or explosion when laser exit is used with flammable materials, solutions or gases 
or in an environment full of oxygen. Avoid using flammable or oxidizable anesthetic gases, such as nitrous 
oxide (N

2

0). Due to high temperatures, some materials, like cotton, when saturated with oxygen, may be 

ignited. Adhesive solvents and flammable solutions used for cleaning and disinfection should be evaporated 

Summary of Contents for Therapy Plus

Page 1: ...User Manual Therapy Plus ENG...

Page 2: ...IN n 4 2012 established by ANVISA provides the documents for download at the following electronic address www dmcgroup com br in the USE INSTRUCTIONS page Check the revision of the User Manual and ANV...

Page 3: ...diabetic ulcers cutaneous ulcers of different etiologies venous stasis ulcer skin scores vascular microcurrent diabetes post surgery contact dermatitis burns pressure ulcer diabetic ulcer and varicos...

Page 4: ...lure Hematologic neoplasms tumors of hematopoietic and lymphoid tissues Uncontrolled glaucoma Pregnant patients Do not apply the laser on tattoos Aesthetic phototherapy On pregnant uterus Neoplasm in...

Page 5: ...nt Output current Continuous current Input power 25 VA Working mode Continuous Voltage frequency 50 60 Hz Protection against shock Internally Powered Protection degree regarding water and solid object...

Page 6: ...ver direct it for any person except he she is on treatment Bright surfaces may reflect laser light towards the eyes Never radiate tumor processes directly the laser can stimulate them Never radiate an...

Page 7: ...heck the integrity of the delivery system If the target beam is not present in the distal end of the delivery system if its integrity is reduced or if it seems diffuse this is a possible indication th...

Page 8: ...r transportation Additional parts or accessories may be purchased from the DMC through the following codes Code Description 010130131 Power cable 010990317 Power Supply 050020001 Safety glasses 050020...

Page 9: ...is blinking fast the laser is active and when it is slow the battery is charging C Left button it selects the red or infrared laser D Right button it selects the energy in Joules and the Ilib techniqu...

Page 10: ...ld be used in dental medical or physiotherapy offices and it should be posi tioned over a bench and connected to power outlet according illustrated below 1M The spacer inserted in the handpiece must b...

Page 11: ...figure below Reinstall the lid of the battery compartment Before switching on the equipment for the first time it is necessary to charge the battery There are two ways to charge the battery connectin...

Page 12: ...ain starting from the point of interruption press the switch button or press the right button to start a new application cycle At the end of the applica tioncycle thetimecounterisrestartedautomaticall...

Page 13: ...INSULATION OF ELECTRICAL NETWORK In case of emergency or to clean the equipment power cable plug must be disconnected from outlet CLEANING DISINFECTION Before being cleaned the equipment should be di...

Page 14: ...iece shall not be broken The optical fiber shall not be damaged Do not try to disassemble the handpiece or its support In case of problem in the equipment operation contact the DMC technical assistanc...

Page 15: ...e risks the customer should discard the equipment according to the local legislation The Therapy Plus battery contains lithium and shall not be disposed in common waste LASER SUPPLY SYSTEM The equipme...

Page 16: ...dA d E Maximum Power 0 25 W Area 1 2 x 10 7 m2 E 20 x 105 W m2 CALCULATION OF CORNEA MPE Applied Standard 60825 1 2014 Table A 2 Wavelength 808 nm 2 25 0 2 6 4 18 m W T C C EMPE t T2 C6 66 7 C4 1 64 T...

Page 17: ...usto mer or user shall ensure it will be used in such environment Immunity test Compliance Electromagnetic Environment Standard RF emissions ABNT NBR IEC CISPR 11 Group 1 Compliance Therapy Plus uses...

Page 18: ...ge IEC 61000 4 5 2 kV line to ground 1 kV line to line Compliance Quality of energy supply should be like a hospital or typical commercial environment Voltage drop short interruptions and vol tage var...

Page 19: ...lower than compliance level of each frequency rangeb It may clash the equipment marked with the symbol Note 1 In 80 MHz and 800 MHz the highest frequency range is applied Note 2 These standards can t...

Page 20: ...rding to transmitter frequency m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 2 1 2 2 3 10 3 7 3 7 7 4 100 12 12 23 For transmitters with maximum n...

Page 21: ...equipment which will be connected to the direct current power supply Continuous current I P 20 Protected against 12 5 mm size solid unknown objects and not protected against water Manufacture date Man...

Page 22: ...the equipment is repaired out of warranty period it can be only extended to exchanged components E The most common defects are physical shocks applied to the equipment and in these cases the warranty...

Page 23: ...UIPAMENTOS LTDA Rua Sebasti o de Moraes 831 Jardim Alvorada S o Carlos SP CEP 13562 030 CNPJ 02 827 605 0001 86 ANVISA Registration 80030810157 Tech Responsible Renaldo Massini Jr CREA 0601706815 Tech...

Page 24: ...For more information about our technology and how to use it visit www nupen com br Technical Assistance DMC IMP E EXP DE EQUIP LTDA 55 16 2107 2323 www dmcgroup com br...

Reviews: