Therapy Plus |
User’s Manual
Features
Specifications
Manufactured and tested according to:
IEC 60601-1, IEC 60601-1-2, IEC 62034, IEC 60601-
2-22 and IEC 60685-1.
Software version
12
Use temperature
10 °C - 30 °C
SAFETY - IMPORTANT PRECAUTIONS
• Please read the whole manual before using the equipment.
• Attention - Using other controls, setting or running not specified in this manual may result in the
exposure to hazardous radiation.
• Attention - Laser smoke may present living tissue particles.
• As the equipment will be used by qualified professional, an additional specific training for its using
is not required. However, the reading of whole manual before its using is recommended.
• If the equipment is not used for some time, the battery shall be removed.
• The replacement of the battery by a personnel with inadequate training might result in a danger
(such as excessive temperature, fire or explosion).
• Laser lights may cause eye injuries. People who are on site should protect themselves against laser
lights emission. Therapy Plus supplies three green glasses (one for operator, the other for assistant
and the other one for patient). Only glasses supplied by DMC should be used with the equipment.
• Never look directly to laser light. Never direct it for any person, except he/she is on treatment.
• Bright surfaces may reflect laser light towards the eyes.
• Never radiate tumor processes directly, the laser can stimulate them.
• Never radiate an undiagnosed injury.
• Only skilled professionals should handle the equipment. An inappropriate use may bring irrevers-
ible injuries.
• Only components mentioned in this manual should be used together with equipment.
• This equipment cannot be used with power cable not supplied by DMC because it may provoke an
increase of emission or decrease of immunity of equipment.
• To avoid risks of shocks, the equipment should be connected only in grounded outlets.
• Do not connect power cable in difficult-to-access outlets.
• Disconnect the power cable from outlet by plug.
• No change should be made in the equipment.
• Do not obstruct equipment air input/output.
• Do not apply any protector film on handpiece in order to obstruct air input/output.
• Do not install the equipment directly on sunlight, dusty places or mechanical vibration.
• There are risks of fire and/or explosion when laser exit is used with flammable materials, solutions or gases
or in an environment full of oxygen. Avoid using flammable or oxidizable anesthetic gases, such as nitrous
oxide (N
2
0). Due to high temperatures, some materials, like cotton, when saturated with oxygen, may be
ignited. Adhesive solvents and flammable solutions used for cleaning and disinfection should be evaporated