20
Therapy Plus |
User’s Manual
Minimal distances recommended for portable and mobile RF communication equipment and Therapy Plus.
Therapy Plus is designated to be used in electromagnetic environment in which radiated RF disturbances are controlled.
Therapy Plus’ customer or user may help avoiding electromagnetic interference keeping a minimal distance between por-
table or mobile RF communication equipment (transmitters) and Therapy Plus, as recommended below, according to the
maximum output power of communication equipment.
Maximum nominal
output power of trans-
mitter (W)
Distance according to transmitter frequency (m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1,2
1,2
2,3
10
3,7
3,7
7,4
100
12
12
23
For transmitters with maximum nominal power not listed above, the recommended distance (in meters [m]) can be deter-
mined by equation applicable to transmitter frequency.
Note 1:
In 80 MHz to 800 MHz, the distance for highest frequency range is applied.
Note 2:
These standards can’t be applied at all situations. Electromagnetic propagation is affected by absorption and
reflection of structures, objects and people.
Table 4: Recommendations for distances between equipment and RF emission source based on Table
206 – 60601-1-2, 2007.
USED SYMBOLS
Laser radiation
Attention
General symbol for warning
Single use
See user’s manual