14
Therapy Plus |
User’s Manual
PREVENTIVE MAINTENANCE
The equipment should be gauged anually by DMC. DMC doesn’t have any responsibility regarding
safety of equipment if calibration is not performed.
All services, including technical assistance, exchanges, repair, calibration, etc., should be performed
only by DMC. DMC doesn’t provide circuit diagrams, component lists, descriptions, instructions for
calibration and gauging for unqualified people.
If maintenance or other type of technical assistance is performed by unauthorized staff, DMC doesn’t
have any responsibility regarding safety of equipment.
Before using the equipment, check the state of the handpiece and the power cords. The cords shall not
be broken, twisted, creased, with their external protection covers cut or internal conductors exposed. The
plastic housing of the handpiece shall not be broken. The optical fiber shall not be damaged.
Do not try to disassemble the handpiece or its support. In case of problem in the equipment
operation, contact the DMC technical assistance.
STORAGE AND TRANSPORTATION
The equipment should be stored in an environment without dust, avoid exposing it to direct sunlight
and keep it away from chemical products and cleaning products.
It should be stored, transported and used in the following environmental conditions:
• Temperature: 10ºC to 40ºC.
• Humidity: 30% to 75%.
• Atmospheric pressure: 700 hPa to 1060 hPa.
Avoid mechanical shocks, blows and vibration in the equipment
TROUBLES AND POSSIBLE SOLUTIONS
Trouble
Possible solutions
The equipment doesn’t turn on.
- Check if battery is well connected.
- Check if outlet is energized and in good condition.
- Recharge the battery.
- Call for DMC’s technical assistance.
The battery charge is exhausted quickly
- Check the battery charge level (see item “USING”)
and recharge it. In case of the charge level does not
change upon charging, contact DMC.