Warning!
Patient hazard –
Before therapy
, a
test run
consisting
of several exercise cycles must be com-
pleted without the patient. Then repeat
the test run with the patient and check
that the movement does not cause any
pain.
Note: See also 2.2 and 2.3, as well as
page VI/VII.
Note!
Programming is only possible when the
patient's chip card has been inserted.
For information about therapy param-
eters and programming of special func-
tions, please refer to sections 5.1 to 5.3.
See section 5.4 for programming
examples.
Important!
The programming unit
ARTROMOT
®
-E2/
-E2 compact "Graphics"
can be connected
to all products of the
ARTROMOT
®
family of elbow CPM machines
.
It is possible to exchange chip cards be-
tween the "Text" and "Graphics" versions.
Please note:
1. If a "Graphics" programming unit is
operating with a formatted "Text"
chip card
− the
selected language
of the
"Text" version remains activated;
it is irrelevant for the "Graphics"
version.
− the
reverse on load
setting saved
in the "Text" version is automati-
cally used for both motors.
− the
key lock
is disabled.
2. If a “
Text
” programming unit is op-
erating with a formatted “Graphics”
chip card
− the
selected language
automati-
cally changes to the default set-
ting: German.
− the
reverse on load
setting saved
for motor A in the "Graphics" ver-
sion is automatically used for both
motors.
− a
key lock
set in the "Graphics"
version will remain active, only
when no changes were made with
the "Text" version; the key lock
function is
not an option with the
"Text" version.