13.1 GUIDANCE AND MANUFACTURER'S
DECLARATION - ELECTROMAGNETIC EMISSION -
FOR ALL EQUIPMENT AND SYSTEMS
Guidance and manufacturer's declaration
- Electromagnetic immunity
The Infrared Thermometer SLS-IRCT30 is intended for use
in the electromagnetic environment specified below. The
customer or the user of the Infrared Thermometer SLS-IRCT30
should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Compliance
Group 1
Class B
Electromagnetic
environment -
guidance
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
Operating
Environment
Temperature: 10°C–40°C (50°F–104°F)
Humidity: 15%–95% RH, non-
condensing
Atmospheric pressure: 86–106 kPa
Storage and
Transportation
Temperature: -20°C to 55°C
(-4°F–131°F)
Humidity: 0–95% RH, non-condensing
Atmospheric pressure: 50–106 kPa
The infrared thermometer has been tested and conforms to
the standard ASTM E1965-98. ASTM laboratory accuracy
requirements in the display range of 98°F to102°F (37°C-39°C)
for skin IR thermometers is ±0.5°F (±0.3°C). Note that
for mercury-in-glass and electronic thermometers, the
requirement per ASTM Standards E667-86 and E1112-86 is
±0.2°F (±0.1°C).