22
13.4 RECOMMENDED SEPARATION DISTANCES
BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE
EQUIPMENT OR SYSTEM - FOR EQUIPMENT AND
SYSTEMS THAT ARE NOT LIFE SUPPORTING
NOTE:
• At 80 MHz and 800 MHz, the separation distance
for the higher frequency range applies.
• These guidelines may not apply in all situations.
Electromagnetic propagation is affected by
absorption and reflection from structures, objects
and people.
The Infrared Thermometer SLS-IRCT30 is intended for use
in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of
the Infrared Thermometer SLS-IRCT30 can help prevent
electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications
equipment (transmitters) and the Infrared Thermometer
SLS-IRCT30 as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power of
transmitter W.
Separation distance according to
frequency of transmitter.
80 MHz to 800
MHz
800 MHz to 2.5
GHz
0.01
0.12
0.23
0.1
0.38
0.73
1
1.2
2.3
10
3.8
7.3
100
12
23
For transmitters rated at a maximum output power not listed
above, the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the
transmitter manufacturer.