Direct Supply ZERO OPENING ASSIST Owner'S Manual Download Page 3

3

1-800-634-7328    DirectSupply.com

Important Warnings

 

WARNING:

 Risk of serious injury or death. This 

product is intended only for use with beds that are 

capable of supporting assist devices via mounting holes 

in the bed’s head deck. This assist device is designed for 

use only on beds with assist device mounting holes which 

align with the slots provided on the assist’s mounting 

plate. Do not modify the assist device or your bed if the 

holes do not align. Do not use this device with any other 

model or brand of bed. Use of this product on any bed 

it was not designed for could result in an unproven or 

unsafe configuration, potentially resulting in serious injury 

or death. For questions regarding compatibility, contact 

your Direct Supply account manager.

 

WARNING:

 This assist device is designed to fit 

multiple models of beds and can be configured to pivot to 

various positions. It may be possible to install this device 

in a manner which could result in entrapment concerns, 

resident injury and/or damage to your bed.

 

WARNING:

 This assist device is designed to pivot to 

four positions, enabling the device to be mounted to either 

side of a bed. Failure to use the pivot-lock system will allow 

the device to pivot toward the floor, potentially creating a 

crush hazard and/or a risk of damage to your bed.

 

WARNING:

 This manual covers the Direct Supply 

Zero Opening Assist Device. For warnings and 

instructions related to the bed itself, please see your 

bed’s Owner’s Manual. Be sure to read and follow all 

instructions and warnings in both manuals.

Bed System Entrapment Information 

In April 1999, the U.S. Food and Drug Administration 

(FDA) in partnership with representatives from the 

hospital and Post-Acute bed industry, national healthcare 

organizations, resident and patient advocacy groups, and 

other federal agencies formed the Hospital Bed Safety 

Workgroup (HBSW). The workgroup’s goal is to improve 

the safety of bed frames for residents and patients 

in all healthcare settings who are most vulnerable to 

the risk of entrapment. The efforts of the FDA and the 

HBSW culminated in the FDA’s release of recommended 

guidelines intended to reduce the risk of entrapment, 

including dimensional limits for critical gaps and spaces 

between bed system components and clinical guidance 

for assessment and implementation of bed assist rails in 

various healthcare settings. Entrapment zones involve the 

relationship of bed components often directly assembled 

by the healthcare facility rather than the manufacturer. 

Therefore, compliance is the responsibility of the facility. 

 

WARNING:

 Only use the included hardware to attach 

this assist device to a bed. Do not substitute hardware.

 

WARNING:

 Risk of serious injury or death. Use 

a properly sized mattress in order to minimize the 

gap between the side of the mattress and the assist 

device. This gap must be small enough to prevent a 

resident from getting his or her head or neck caught 

in this location. Make sure raising or lowering the 

bed, or articulating the sleep surface, does not create 

hazardous gaps. Failure to do so could result in serious 

injury or death.

 

WARNING:

 Risk of serious injury or death. Do not 

use this assist device if any openings within the assist 

device body will allow a resident to get his or her head 

or neck lodged within these openings. Failure to do so 

could result in serious injury or death.

 

WARNING:

 Risk of serious injury or death. Do not 

use this assist device until you have verified it is locked 

in place. Injury to resident or caregiver may result if 

this procedure is not followed.

 

 

WARNING:

 An optimal bed system assessment 

should be conducted on each resident by a qualified 

clinician or medical provider to ensure maximum safety 

of the resident. The assessment should be conducted 

within the context of, and in compliance with, the 

state and federal guidelines related to the use of 

restraints and bed system entrapment guidance, 

including the 

Clinical Guidance for the Assessment 

and Implementation of Side Rails

 published by the 

Hospital Bed Safety Workgroup of the U.S. Food and 

Drug Administration. 

 

WARNING:

 In the event worn or broken parts 

are identified, remove the device from service until 

repairs are made. Do not use substitute parts. Use 

only approved replacement parts. Non-approved 

replacement parts could cause entrapment, personal 

injury and damage to assist devices.

NOTE:

 The assist device is intended for use as an 

aid in entering or exiting the bed sleep area, as well 

as a stable handhold during self-positioning within 

the bed sleep area. The devices may deform or 

break when subjected to excessive side pressure. 

Do not exert excessive side pressure on the assist 

devices. Do not use them as push handles for 

moving the bed. These activities could result in 

personal injury and damage to the assist devices 

and/or your bed.

This assist device is only one part of your healthcare bed 

system. Proper combinations of bed, mattress, head/

foot panels and assist rails are needed to minimize the 

risk of entrapment. For more information, contact your 

Direct Supply account manager.

 

WARNING:

 Do not permit residents to climb over 

the device. 

 

WARNING:

 Some states may prohibit assist devices 

or consider assist devices to be resident restraints in all 

or some instances. Always assess whether this assist 

device may meet the definition of a restraint on a  

case-by-case basis and document the need for any 

resident restraint. Contact your state surveyor for 

further guidance regarding assist devices and restraints.

Warnings

Summary of Contents for ZERO OPENING ASSIST

Page 1: ...t Device from Direct Supply Equipment Furnishings Please read this entire guide carefully and keep it for future reference This guide provides instructions warnings warranty information and other impo...

Page 2: ...via holes in the bed s head deck See the installation section of this manual for information about device setup and ensuring compatibility and safety This device is not a bedrail and is not intended...

Page 3: ...substitute hardware WARNING Risk of serious injury or death Use a properly sized mattress in order to minimize the gap between the side of the mattress and the assist device This gap must be small en...

Page 4: ...ing either direction right or left If setting up devices on both sides of a bed align slots so assists on opposite sides of the bed are oriented as mirror images of each other Additionally check the o...

Page 5: ...r hole D or hole B will prevent pivoting in the opposite direction of screw placement Fully insert but do not tighten Insert one set screw into hole B and one set screw into hole D using a 3 mm Allen...

Page 6: ...not substitute hardware Safe installation requires use of all provided attachment hardware Installation cont 6 To prevent rapid height adjustment or removal of the assist use the included nut and bolt...

Page 7: ...djust the pivot lock system to prevent pivoting toward the head of the bed see step 4 for guidance on how to set up your pivot lock system c Review the overall bed system for compliance to FDA entrapm...

Page 8: ...vice left or right 90 degrees NOTE Do not rotate assist device more than 90 degrees 2 Rotate handle to lock into position 3 Adjust handle height by removing the release pin or bolt moving the handle t...

Page 9: ...r excessive wear 3 Visually examine all welds for cracks 4 Tighten attachment bolts and center pivot screw 5 Inspect all bolts and fasteners Do not overtighten bolts at pivot points Cleaning Clean wit...

Page 10: ...the applicable limited warranty period identified below whichever is earlier Warranty Period Parts Only Anticipated Usable Device Life Direct Supply Zero Opening Assist Device 3 years 3 years Anticip...

Page 11: ...Zero Opening Assist and enjoy outrageous customer service If you have any questions about the zero opening assist you have purchased or would like to request warranty service please contact Direct Su...

Page 12: ...35 West Champions Way Milwaukee WI 53223 2020 Direct Supply Manufacturing Inc All Rights Reserved Direct Supply Panacea and all associated logos are the registered trademarks of Direct Supply Inc used...