3. Changing Seat Height and Tilt Angle

Normal up/down; Rapid up/down. NOTE! The seat has a soft start for 
around 5 sec. before the correct speed is achieved. Make sure the footrest 
does not touch the floor when the seat is at lower levels – this can happen if 
the footrest is set for a tall user.

Tilting forwards/rearwards: The chair can be tilted 5 degrees forwards* 
and 30 degrees rearwards. When tilting, the chair will always stop at the 
horizontal position. Thus to continue tilting forwards or rearwards, the 
forwards/rearwards button must be pressed again. When tilting forwards, 
make sure the footrest does not touch the floor or frame – this can happen if 
the footrest is set for a tall user.

(*In certain markets, the chair can be tilted further to 6 degrees forwards.
Note: Crush risk if this is done.).

Emergency lowering: if battery charge is insufficient for chair 
operation or if the proximity sensor on the height adjustment   

 

 

actuator is defective, the Emergency lowering button must be used. 
Upon activation, the chair will revert to the horizontal position and 
then to the lowest seat height. Exercise caution when activating as    

 

the system safety sensors are bypassed. In particular, make sure the 
chair proceeds evenly to its minimum height to prevent skewing.

Service indicator – the lamp will flash at 4 sec. intervals when actual 
operating time reaches 11 months from the last reset. After 12 months, the 
lamp will remain on. This indicates the need for scheduled maintenance 
(page 8). Resetting the service indicator – see technical manual.

Battery indicator – shows current battery status; 3 bars indicate full charge. 
When a single bar shows, the chair should be charged that same day. When 
a red bar shows, stop using the chair and charge the battery. Make a habit 
of always charging the battery if at the end of the day it has dropped to level 
1 (one green bar on the battery indicator). If the Hygiene Chair will not be 
used for a few days, we recommend switching off the battery by means of 
the Off button located on the battery. 

NOTE: The battery can be damaged if not charged at least every 14 days in 
the On position or at least every three months in the Off position.

The remote control enters sleep mode 30 sec. after the last activation.

Snabb
ned

Ned

Bakåt

Service- 
indikator

Snabb

upp

Upp

Framåt

Nöd- 

sänkning

Batteri-

indikator

Summary of Contents for 56-270

Page 1: ...EASY SoftBack User Manual Technical Manual Service Manual DIRECTHEALTH C ARE GR OUP COM Manual no 805 rev 1 210609 SWL 150 kg 330 lbs 56 270 56 270 CS 56 270 100 CS 56 270 CS UK 56 270 LOW CS...

Page 2: ...11 23 Dimensions 13 24 Information on EMC 14 25 Further Information om EMC Advanced 14 26 Parts List 17 27 Spare Parts 17 28 Accessories 19 29 Table of Symbols 19 EASY is a well designed Hygiene Chair...

Page 3: ...ights up after around 11 months operation as a reminder regarding the scheduled service that must be carried out at least once per year Information about how to reset the service indicator can be foun...

Page 4: ...se after every folding disassembly and after any transport Check the battery status prior to and after each use It is important to never leave the user alone during the transfer Only transfer the user...

Page 5: ...ion or at least every three months in the Off position Make sure protective covers are always put back in place after cleaning activities Always replace the Hygiene Chair s electrical components or ca...

Page 6: ...ontal position and then to the lowest seat height Exercise caution when activating as the system safety sensors are bypassed In particular make sure the chair proceeds evenly to its minimum height to...

Page 7: ...y get in and damage the remote control 5 Battery On Off The battery is located under a protective cover at the base of the chair The battery has On Off buttons that are reached through an opening at t...

Page 8: ...h the one on the righthand front side features a directional lock and the other 3 are fitted with full brakes When the directional lock is activated wheels cannot be turned This is particularly useful...

Page 9: ...juries make sure protective covers are always put back 17 Troubleshooting when the Chair is Inoperable NOTE Chair cannot be operated above 65 cm seat height if the chair is tilted forward nor can the...

Page 10: ...eries must be removed and recycled as electrical components Ask your local recycling station for further information about how the different materials in the product metals plastics electronics should...

Page 11: ...the pillars with detergent disinfectants commonly used in healthcare alcohol based Make sure the protective cover is installed at the bottom of each pillar Special checks at least once per year 1 Armr...

Page 12: ...ng temperature 5 45 C 5 45 C Battery TBB6 Voltage 24 V Capacity 5 Ah Type Pb Protection class IP66W Battery charger TP9 Voltage in 100 240V 50 60 Hz Voltage 29V Control box TC14 Processor MCU 3 motors...

Page 13: ...ntrol box 2 Actuators up down left 5 Remote control 3 Actuators up down right 6 Battery 4 Actuators tilt 7 Connects to outgoing connector on control box 7 Connected to spindle cable 7 23 Dimensions Al...

Page 14: ...proval Avoid placing using the chair less than 0 35m from other medical equipment and or mobile phones or other transmitting equipment This is because there is then a risk that the main functions of t...

Page 15: ...e 1 kV differential mode n a for common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply...

Page 16: ...be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range...

Page 17: ...rest left right 11 58 296 1 Footrest 12 58 281 1 Bottom frame 13 57 049 RF Wheel with bolt fully braked 13 57 040 RF Wheel with bolt directional lock and fully braked 14 58 360 Cover 15 57 793 End plu...

Page 18: ...EASY SOFTBACK 18 DIRECTHEALTHCAREGROUP COM...

Page 19: ...232 Plug soft seat 56 234 Comfort seat back opening green Easy 56 259 L Calf rest left Easy 56 259 R Calf rest right Easy 56 260 L Footrest left Easy 56 260 R Footrest right Easy This product complie...

Page 20: ...tan 35 SE 164 40 Kista Sweden Tel 46 0 8 557 62 200 info directhealthcaregroup com www directhealthcaregroup com Medical Device Class I The product complies with the re quirements of the Medical Devic...

Reviews: