EASY SOFTBACK

21. Maintenance

The chair is made from stainless steel/powder coated tubing and aluminium.
Exposed, unpainted parts are made of aluminium or stainless steel. Make sure the 
product feels firm and stable – no fasteners may be loose. Damaged or defective 
parts should be replaced as soon as possible.

Cleaning, general

•  Clean the chair by wiping with regular cleaning agent (NOT Pentanol based) or 

use a washer/autoclave (max 85 C).

•  Remove the seat and clean it separately or use a washer/autoclave (max 85 C). 

NOTE: The remote control/control box (or the connector/
female connector if the remote control is removed before 
washing) must be protected from high pressure water.
Use a plastic bag or similar that can withstand washer water pressure.  
Place the remote control/control box or the connector/female 
connector so that running watercannot force its way in.

•  Clean the pillars with detergent/disinfectants 

commonly used in healthcare (alcohol-based).

•  Make sure the protective cover is installed at the bottom of 

each pillar.

Special checks (at least once per year)

1.  Armrest – check that slip protection/wrist protection 

are intact and secured.

2.  Pillars (electric actuator/raise-lower) + electric actuator/ 

tilt: check the function of the remote control and that  
the electric actuator attachment devices are not loose.

3.  Seat/stopped seat – whole and without cracks. No sharp edges may be 

present.

4.  Footrest platforms – whole and adjustable in the intended manner. Check that 

the clips are whole and properly fastened. Heel restraints must be in position at 
the back of each footrest platform.

5.  Wheels – may not be stiff or loose. Also check that the brakes work.

6.  Back strap; the back strap must be intact and may not have any holes or other 

damage to the belts and anchors.

6

1

3

2

4

5

22. Technical Specifications

Material

Aluminium
Plastic (PP, PP/GS, PA, PA/GS, TPE, Silicone)
Textile Dartex
Foam (PUR)
Steel (Stainless)
Frame / stand Powder coating (polyester)

Specification power system

Summary of Contents for 56-270

Page 1: ...EASY SoftBack User Manual Technical Manual Service Manual DIRECTHEALTH C ARE GR OUP COM Manual no 805 rev 1 210609 SWL 150 kg 330 lbs 56 270 56 270 CS 56 270 100 CS 56 270 CS UK 56 270 LOW CS...

Page 2: ...11 23 Dimensions 13 24 Information on EMC 14 25 Further Information om EMC Advanced 14 26 Parts List 17 27 Spare Parts 17 28 Accessories 19 29 Table of Symbols 19 EASY is a well designed Hygiene Chair...

Page 3: ...ights up after around 11 months operation as a reminder regarding the scheduled service that must be carried out at least once per year Information about how to reset the service indicator can be foun...

Page 4: ...se after every folding disassembly and after any transport Check the battery status prior to and after each use It is important to never leave the user alone during the transfer Only transfer the user...

Page 5: ...ion or at least every three months in the Off position Make sure protective covers are always put back in place after cleaning activities Always replace the Hygiene Chair s electrical components or ca...

Page 6: ...ontal position and then to the lowest seat height Exercise caution when activating as the system safety sensors are bypassed In particular make sure the chair proceeds evenly to its minimum height to...

Page 7: ...y get in and damage the remote control 5 Battery On Off The battery is located under a protective cover at the base of the chair The battery has On Off buttons that are reached through an opening at t...

Page 8: ...h the one on the righthand front side features a directional lock and the other 3 are fitted with full brakes When the directional lock is activated wheels cannot be turned This is particularly useful...

Page 9: ...juries make sure protective covers are always put back 17 Troubleshooting when the Chair is Inoperable NOTE Chair cannot be operated above 65 cm seat height if the chair is tilted forward nor can the...

Page 10: ...eries must be removed and recycled as electrical components Ask your local recycling station for further information about how the different materials in the product metals plastics electronics should...

Page 11: ...the pillars with detergent disinfectants commonly used in healthcare alcohol based Make sure the protective cover is installed at the bottom of each pillar Special checks at least once per year 1 Armr...

Page 12: ...ng temperature 5 45 C 5 45 C Battery TBB6 Voltage 24 V Capacity 5 Ah Type Pb Protection class IP66W Battery charger TP9 Voltage in 100 240V 50 60 Hz Voltage 29V Control box TC14 Processor MCU 3 motors...

Page 13: ...ntrol box 2 Actuators up down left 5 Remote control 3 Actuators up down right 6 Battery 4 Actuators tilt 7 Connects to outgoing connector on control box 7 Connected to spindle cable 7 23 Dimensions Al...

Page 14: ...proval Avoid placing using the chair less than 0 35m from other medical equipment and or mobile phones or other transmitting equipment This is because there is then a risk that the main functions of t...

Page 15: ...e 1 kV differential mode n a for common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply...

Page 16: ...be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range...

Page 17: ...rest left right 11 58 296 1 Footrest 12 58 281 1 Bottom frame 13 57 049 RF Wheel with bolt fully braked 13 57 040 RF Wheel with bolt directional lock and fully braked 14 58 360 Cover 15 57 793 End plu...

Page 18: ...EASY SOFTBACK 18 DIRECTHEALTHCAREGROUP COM...

Page 19: ...232 Plug soft seat 56 234 Comfort seat back opening green Easy 56 259 L Calf rest left Easy 56 259 R Calf rest right Easy 56 260 L Footrest left Easy 56 260 R Footrest right Easy This product complie...

Page 20: ...tan 35 SE 164 40 Kista Sweden Tel 46 0 8 557 62 200 info directhealthcaregroup com www directhealthcaregroup com Medical Device Class I The product complies with the re quirements of the Medical Devic...

Reviews: