UltiMate 3000 Nano and Cap- System Installation and Application
54
•
Select
None
to disable synchronization.
•
When the UltiMate 3000 system includes a DGP-3600 pump:
Select
UM3PUMP _L_STRK
to enable synchronization with the left pump. To enable
synchronization with the right pump, select
UM3PUMP _R_STRK
.
•
When the UltiMate 3000 system includes an LPG-3400 pump:
Select
UM3PUMP _STROKE
to enable synchronization.
Please note:
You can change the default synchronization assignment for a specific
application, using the
PumpDevice
and
SyncWithPump
properties in the
Commands
dialog box under
Sampler
on the control panel or in the
program file (PGM).
To disable synchronization for a specific application, set
SyncWithPump
to
Off
. Use the
PumpDevice
property to select the pump for which
synchronization shall be performed. (The drop-down list contains names
of the pumps installed in the Server Configuration program.)
Changing the settings for these properties from the
Commands
dialog box
does not change the standard synchronization setting on the
Segments
tab
page.
For more information, refer to the
Operating Instructions
for the autosampler and/or pump or
to the Chromeleon
Online Help
.
4.5 Displaying the Degasser-Related Parameters
If your system includes an SRD-3x00 solvent rack, the degasser can be controlled with
Chromeleon via the pump. In this case, the following degasser-related parameters are listed in
the
Commands
dialog box under the pump to which the solvent rack is connected:
Chromeleon
Description
Degasser
Turns the solvent rack degasser on or off.
For pumps operated with an SRD-3000 solvent rack, the setting
should always be
On
. For pumps that are operated without degasser
or with an external third-party degasser, select
Off
.
DegasserVacuum
Indicates the status of the degasser vacuum in the solvent rack
(
OK
or
NotOK
; read-only).
SolventRackLeak
Indicates the status of the leak sensor in the solvent rack
(
Leak
or
NoLeak
; read-only).
Please note:
Dionex recommends always leaving the degasser on.
Turning off the pump automatically turns off the degasser, too.
Summary of Contents for UltiMate 3000
Page 2: ...UltiMate 3000 Nano and Cap System Installation and Application...
Page 10: ...UltiMate 3000 Nano and Cap System Installation and Application 6...
Page 60: ...UltiMate 3000 Nano and Cap System Installation and Application 56...
Page 66: ...UltiMate 3000 Nano and Cap System Installation and Application 62...
Page 72: ...UltiMate 3000 Nano and Cap System Installation and Application 68...