background image

 

IonPac TCC-LP1, TCC-ULP1, and TCC-XLP1 
Product Manual 

Page 7 of 17 

 

 

 

 

 

 

Doc. No. 034973-06 

©2005 DIONEX 

September 2005 

 

3.

 

OPERATION 

 

3.1.

 

Sample Loading 

WARNING

Always use the high pressure pulse damper (Dionex P/N 043945) after the DQP pump or DXP pump 
to ensure the concentrator column does not get exposed to pump pulsations.  It is possible to damage 
the resin when exposed to high pump pulsations.

WARNING

To prevent overloading the TCC-LP1, TCC-ULP1, or TCC-XLP1, and/or loss of sample analytes, 
determine the concentration linearity over the desired analytical concentration range.  See Section 
3.4.1, “Capacity Consideration of Concentrators.”

 
Sample loading can be performed manually with a < 3 mL syringe.  It takes approximately 1 minute to manually inject 3 mL 
of sample through the TCC-LP1. 
 
Alternately, you can use a separate positive displacement pump such as the Dionex DQP pump (P/N 035250).  When using a 
pump, ensure that a high pressure damper is used.  Pump flow rates of approximately 3 mL/min can be used while 
maintaining sample concentration efficiencies high enough to ensure good quantification. 
 

WARNING

The flow direction during the concentration step is critical.  In order to ensure optimal system 
performance, it is recommended that concentration always be performed in a back flush manner.

 
After the sample has been loaded onto the TCC-LP1, TCC-ULP1, or TCC-XLP1 in the direction opposite to the eluent flow, 
it is then “back flushed” with eluent onto the guard and analytical columns (see Figure 4, “Loading the TCC-LP1, TCC-
ULP1, or TCC-XLP1 Column”).  This configuration concentrates the cations in a tight band at the bottom of the TCC 
columns.  When injected, all of the ions are rapidly eluted off of the TCC columns, and onto the guard and analytical 
columns.  If the sample is loaded onto the TCC-LP1, TCC-ULP1, or TCC-XLP1 in the same flow direction as the eluent 
flow, the cations are concentrated at the head of the column rather than at the bottom.  When injected, the cations begin 
chromatographic separation on the concentrator before reaching the guard and analytical columns.  Therefore, the retention 
time of the analytes would be significantly longer than a standard loop injection.  Normally the function of the concentrator is 
to strip the ions of interest from the sample matrix and not to act as an analytical column. 
 
 

Summary of Contents for IonPac TCC Series

Page 1: ...IonPac TCC...

Page 2: ...ptember 2005 PRODUCT MANUAL FOR TRACE CATION CONCENTRATOR TCC TCC LP1 Column Low Pressure 4 x 35 mm P N 063079 TCC ULP1 Column Ultra Low Pressure 5 x 23 mm P N 063783 TCC XLP1 Column Extremely Low Pre...

Page 3: ...acity 10 4 EXAMPLE APPLICATIONS 13 4 1 Concentration in the Manual Mode 13 4 2 Concentration in the Manual Mode 14 5 TROUBLESHOOTING GUIDE 15 5 1 High Backpressure from a Contaminated Inlet Bed Suppor...

Page 4: ...throughout this document to emphasize important points The symbols are shown below along with the purpose of the information SAFETY Safety information can help prevent bodily harm WARNING Warning inf...

Page 5: ...CC ULP1 is less than 45 psi 0 31 MPa at 1 mL min and the TCC XLP1 is less than 30 psi 0 21 MPa at 1 mL min The large resin particle size 20 m in the TCC LP1 TCC ULP1 and TCC XLP1 makes it possible to...

Page 6: ...ional Group Hydro phobicity TCC LP1 4 mm 20 55 260 eq col Carboxylic Acid Very low TCC ULP1 5 mm 20 55 260 eq col Carboxylic Acid Very low TCC XLP1 6 mm 20 55 260 eq col Carboxylic Acid Very low TABLE...

Page 7: ...Column Description The IonPac Trace Cation Concentrator Low Pressure TCC LP1 4x35 mm and IonPac Trace Cation Concentrator Ultra Low Pressure TCC ULP1 5x23 mm columns consist of the following component...

Page 8: ...ugh to ensure good quantification WARNING The flow direction during the concentration step is critical In order to ensure optimal system performance it is recommended that concentration always be perf...

Page 9: ...IONEX September 2005 Figure 3 shows the configuration for sample loading using a Rheodyne valve FIGURE 3 Configuration for Determining Trace Levels of Cations Figure 4 shows the configuration for samp...

Page 10: ...o 8 days Recommended storage vessels are Corning tissue culture flasks The following procedure should be used for storage of g L level samples A Rinse the polystyrene container and cap twice with deio...

Page 11: ...he analytes on the ion exchange sites and result in their elution to waste during the loading process B Conversely qualitative analysis of ions with higher affinities for the resin in the presence of...

Page 12: ...e rate at which the sample is loaded C The pH of the sample D The ionic strength of the sample E The amount and capacity of resin in the column The breakthrough volume is determined as follows A Prepa...

Page 13: ...2005 DIONEX September 2005 FIGURE 6 Linearity Determinations for Concentrator Injection Time min that the simulated sample is pumped through the TCC LP1 TCC ULP1 or TCC XLP1 after the single ion injec...

Page 14: ...thanesulfonic acid Eluent Flow Rate 1 0 mL min Injection Volume 1 0 mL Temperature 35 C SRS Suppressor Cation Self Regenerating Suppressor CSRS ULTRA AutoSuppression Recycle Mode MMS Regenerant 100 mN...

Page 15: ...n Eluent 10 mM Methanesulfonic acid Eluent Flow Rate 1 0 mL min Injection Volume 1 0 mL SRS Suppressor Cation Self Regenerating Suppressor CSRS ULTRA AutoSuppression Recycle Mode or AES Suppressor Cat...

Page 16: ...back on to the column Follow the instructions below to change the bed support assembly using one of the two spare bed support assemblies included in the ship kit provided with the column A Disconnect...

Page 17: ...D Be sure that the eluents are made from chemicals with the recommended purity see Section 3 Operation E Be sure that the deionized water used to prepare the reagents has a specific resistance of 18...

Page 18: ...he gradient pump Direct the effluent from the TCC LP1 TCC ULP1 or TCC XLP1 directly to a waste container C Set the flow rate to 1 mL min D Pump the 1 M HCl 0 1 M KCl solution through the column for 15...

Reviews: