background image

 

IonPac TCC-LP1, TCC-ULP1, and TCC-XLP1 
Product Manual 

Page 14 of 17 

 

 

 

 

 

 

Doc. No. 034973-06 

©2005 DIONEX 

September 2005 

 

4.2.

 

Concentration in the Manual Mode  

The following example demonstrates the advantages of sample concentration with low sensitivity detection versus 
direct injection with high sensitivity detection.  

Sample Loop Volume:  

See Chromatogram  

Concentrator Column: 

TCC-LP1 4 x 35mm 

Analytical Column:  

IonPac CS14 Analytical Column  

Eluent:  

10 mM Methanesulfonic acid  

Eluent Flow Rate:  

1.0 mL/min  

Injection Volume: 

1.0 mL 

SRS Suppressor:  

Cation Self-Regenerating Suppressor, CSRS

®

-ULTRA 

AutoSuppression

 Recycle Mode 

or AES Suppressor 

Cation Atlas

®

 Electrolytic Suppressor, CAES 

or MMS Suppressor 

Cation MicroMembrane Suppressor, CMMS

®

 III 

MMS Regenerant:  

100 mN TBAOH (Use Dionex Cation Regenerant Solution P/N 039602)  

Expected Background Conductivity: 

< 2 µS 

Expected System Operating Back Pressure: 

< 1,300 psi (8.96 MPa) 

 
 

Peaks 

 

  µg/L 

(ppb) 

1.

 

Lithium     

0.5 

2.

 

Sodium    

2.0 

3.

 

Ammonium   

2.5 

4.

 

Potassium 5.0 

5.

 

Magnesium  

2.5 

6.

 

Calcium   

5.0 

 

 

FIGURE 9 

Manual Concentration versus Direct Injection 

Summary of Contents for IonPac TCC Series

Page 1: ...IonPac TCC...

Page 2: ...ptember 2005 PRODUCT MANUAL FOR TRACE CATION CONCENTRATOR TCC TCC LP1 Column Low Pressure 4 x 35 mm P N 063079 TCC ULP1 Column Ultra Low Pressure 5 x 23 mm P N 063783 TCC XLP1 Column Extremely Low Pre...

Page 3: ...acity 10 4 EXAMPLE APPLICATIONS 13 4 1 Concentration in the Manual Mode 13 4 2 Concentration in the Manual Mode 14 5 TROUBLESHOOTING GUIDE 15 5 1 High Backpressure from a Contaminated Inlet Bed Suppor...

Page 4: ...throughout this document to emphasize important points The symbols are shown below along with the purpose of the information SAFETY Safety information can help prevent bodily harm WARNING Warning inf...

Page 5: ...CC ULP1 is less than 45 psi 0 31 MPa at 1 mL min and the TCC XLP1 is less than 30 psi 0 21 MPa at 1 mL min The large resin particle size 20 m in the TCC LP1 TCC ULP1 and TCC XLP1 makes it possible to...

Page 6: ...ional Group Hydro phobicity TCC LP1 4 mm 20 55 260 eq col Carboxylic Acid Very low TCC ULP1 5 mm 20 55 260 eq col Carboxylic Acid Very low TCC XLP1 6 mm 20 55 260 eq col Carboxylic Acid Very low TABLE...

Page 7: ...Column Description The IonPac Trace Cation Concentrator Low Pressure TCC LP1 4x35 mm and IonPac Trace Cation Concentrator Ultra Low Pressure TCC ULP1 5x23 mm columns consist of the following component...

Page 8: ...ugh to ensure good quantification WARNING The flow direction during the concentration step is critical In order to ensure optimal system performance it is recommended that concentration always be perf...

Page 9: ...IONEX September 2005 Figure 3 shows the configuration for sample loading using a Rheodyne valve FIGURE 3 Configuration for Determining Trace Levels of Cations Figure 4 shows the configuration for samp...

Page 10: ...o 8 days Recommended storage vessels are Corning tissue culture flasks The following procedure should be used for storage of g L level samples A Rinse the polystyrene container and cap twice with deio...

Page 11: ...he analytes on the ion exchange sites and result in their elution to waste during the loading process B Conversely qualitative analysis of ions with higher affinities for the resin in the presence of...

Page 12: ...e rate at which the sample is loaded C The pH of the sample D The ionic strength of the sample E The amount and capacity of resin in the column The breakthrough volume is determined as follows A Prepa...

Page 13: ...2005 DIONEX September 2005 FIGURE 6 Linearity Determinations for Concentrator Injection Time min that the simulated sample is pumped through the TCC LP1 TCC ULP1 or TCC XLP1 after the single ion injec...

Page 14: ...thanesulfonic acid Eluent Flow Rate 1 0 mL min Injection Volume 1 0 mL Temperature 35 C SRS Suppressor Cation Self Regenerating Suppressor CSRS ULTRA AutoSuppression Recycle Mode MMS Regenerant 100 mN...

Page 15: ...n Eluent 10 mM Methanesulfonic acid Eluent Flow Rate 1 0 mL min Injection Volume 1 0 mL SRS Suppressor Cation Self Regenerating Suppressor CSRS ULTRA AutoSuppression Recycle Mode or AES Suppressor Cat...

Page 16: ...back on to the column Follow the instructions below to change the bed support assembly using one of the two spare bed support assemblies included in the ship kit provided with the column A Disconnect...

Page 17: ...D Be sure that the eluents are made from chemicals with the recommended purity see Section 3 Operation E Be sure that the deionized water used to prepare the reagents has a specific resistance of 18...

Page 18: ...he gradient pump Direct the effluent from the TCC LP1 TCC ULP1 or TCC XLP1 directly to a waste container C Set the flow rate to 1 mL min D Pump the 1 M HCl 0 1 M KCl solution through the column for 15...

Reviews: