26
Optimized layout for A4 print.
888888
1
1
2
2
3
3
4
4
Calib
6
TYPE
Setup
Ind.Ch
ConfiG
nChan
Chan
A
C
2
1
1
2
DECI
Cel.sen
U.M.
Cel.Cap
DIV
Dead.ld
RANGE1
kno.wGt
theo.CA
7
888888
MENU ACCESS:
SAVING THE PARAMETERS:
Press the
key during the start-
up procedure.
Press the
key several times, until the display shows
SAVE?
. Press the
key to confirm.
Complete menu on pages
24 - 25
USER_MAN_ENG_DGT100_V8
Independent channels / transm
CALIBRATION PROCEDURE:
1.
Select mode of use
ind.Ch
.
2.
Set the number of connected scales (from
1
to
4
).
3.
Select the scale to be calibrated (from
1
to
4
).
4.
Set the calibration parameters:
deCi
= Number of decimals.
u.m
.
= Unit of measurement (
kG
,
G
,
t
,
lb
).
div
= Minimum division.
ranGe1
= Maximum range.
5.
Set the cell data:
sen.Cel
= Cell sensitivity (given by the sum of the value of each cell).
Cel.Cap
= Total capacity of the cells (given by the sum of the value of each cell).
6.
Enter the weight value of the structure in the
dead.ld
parameter. If you do not know this value, enter “0”.
7.
If the structure contains a quantity of material whose weight value is known (e.g. full silo), enter this value in the
KNo.WGT
parameter.
8.
Application of theoretical calibration:
Press the key to exit the calibration menu. The display shows
th.Cal?
. Press the key to confirm the use of the theoretical calibration,
or the key to cancel.
9.
If
transm
mode has been selected, repeat the procedure from point
3
for each scale to be calibrated.