CALIBRATION PROCEDURE:
1.
Select mode of use
dep.Ch
.
2.
Set the number of channels used (from
1
to
4
).
3.
Set the calibration parameters:
deCi
= Number of decimals.
u.m
.
= Unit of measurement (
kG
,
G
,
t
,
lb
).
div
= Minimum division.
ranGe1
= Maximum range.
4.
Set the cell data:
sen.Cel
= Cell sensitivity (given by the sum of the mV/V value of each cell).
Cel.Cap
= Total capacity of the cells (given by the sum of the capacities of each cell).
5.
Enter the weight value of the structure in the
dead.ld
parameter. If you do not know this value, enter “0”.
6.
If the structure contains a quantity of material whose weight value is known (e.g. full silo), enter this value in the
KNo.WGT
parameter.
7.
Application of theoretical calibration:
Press the key to exit the calibration menu. The display shows
th.Cal?
. Press the key to confirm the use of the theoretical calibration,
or the key to cancel.
Theoretical calibration
Dependent channels