27
will be displayed first. The upper row of the display panel displays the instrument
model and software version number, and the lower row displays the version date. Then,
the internal self-test of the instrument starts. If any error is found, the error code
number will be displayed. Multiple errors will be displayed for a certain time in turn,
and then the normal working cycle will be started. If the process number is set,The
correspond workflow will be automatically started.
Inside this instrument, there is a button cell to save the working state before
power failure, and these states will be recalled after power-up. For workflow, after
power-on, the process controller will perform a power-on trigger function, and if the
process has corresponding operations, it will be executed.
After the power supply is turned on, if all the following conditions can be met,
the instrument will perform a weight zero setting, which is to set zero for startup:
(1) The workflow is not in the control state of feeding or discharging.
(2) The weight can collect stable data within 6 seconds after power-on.
(3) The weight value is within the range of starting and zeroing (see F parameter
table F04).
8.2.1 Boot failure code
During the self-test of the instrument after power-on, the following error codes
may be displayed:
serial
number
Code
display
Fault meaning
processing method
1
[
EPr ,, 1
]
Internal
RAM
failure
Repair in factory
2
[
EPr ,, 2
]
Power
failure
detection fault
Check whether the input voltage is
normal.
3
[
EPr ,, 3
]
Loss of internal
RAM data
Check whether the button cell on the
motherboard is dead.
4
[
Epr 4
]
Loss of internal
ROM data
Repair in factory
5
[
EPr 5
]
Internal program
data error
Repair in factory
6
[
Epr, 20
]
Internal
clock
error
Check whether the button cell on the
motherboard is dead.
Table 8.2 Boot Fault Display Codes
8.2.2 Fault codes in normal operation
During the use of the instrument, the following fault codes may also be displayed:
serial
number
Code
display
Fault meaning
processing method
1
[
Err, 03
]
Weight overload
Check the load, sensor, or calibration
coefficient on the scale.
2
[
Er1, 06
]
1# sensor failure
Check 1# sensor, main line, terminal
connection line and junction box.