Installation Guide
33 of 33
Revision 2.00 of 04/11/2009
Use the following evaluation chart.
Appendix A
Suggested procedure to Set sample LOW code .
1) Prepare 3 tubes with 1,5 mL.of blodd
2) Enable the reading view page ( insert the password 131313 like when you enter in the
service menu), before to press START.( " SM MODE ").
3) after the mixing of the samples, check the values for READ1
The value for SL_THRESHOLD parameter will be the average obtained value from vescube
minus 5 units .
E.g 131, 128, 133. set in vescube.ini file ,SL_THRESHOLD = 125.
Michele Meloni
Francesco Cocola
ALL SAMPLES IN THE LIMITS
+/-12
YES
==============
TEST PASSED
NO
==============
PRESENCE OF (*) OR DIFFERENT ERROR
CODE
NO
WITHOUT ERROR CODE
REMOVE RESULTS WITH (*) OR ERROR
CODE FROM THE GRAPH
ALL SAMPLES IN THE LIMITS
+/-12
TEST FAILED ! PERFORM
MORE CONTROLS OR SEND lOg
FILE TO T.A.
YES
NO