Installation Guide
14 of 33
Revision 2.00 of 04/11/2009
The internal barcode reader, placed inside the Preparor Module, is:
•
mechanically set to read labels attached to the tube at least mm.3 above the
the start of bottom roundness (Fig. 2,
).
•
programmed to read barcodes placed at 90° degrees compared to the reading
rays, that is the barcode must be placed to cross horizontally the tubes vertical
axis (Fig. 2,
).
The reader nevertheless can correctly read sloping barcodes, the slope must be within
“-5" and “+5” degrees (Fig. 2,
).
Fig.2 Labelling: compatible heights with Ves-Matic Cube 80
The reading group sensors are able to read correctly the sedimentation rate inside
each sample, following the reading axis, passing through a maximum of three label
layers: thus are allowed only two labels attached to the tube which must be staggered
at least of 90° degrees (Fig. n°3).
Fig.3 Labelling: Correct labels placing on sample tube