Installation Guide
Revision 2.00 of 04/11/2009
Page 12 of 33
8.
Now from the service menu, in the “Procedures” section (located in the bottom-right
corner of the display) end select the “TEST LOAD SAMPLE FROM RACK” end push exec.
•
If the clamp don’t correctly goes on the tube you can change the parameter “CLAMPH
RACK POS1” .
•
If the clamp don’t correctly pick-up the tube from the rack(underneath the cap) you
can change the parameter “LOAD CLAMPV RACK” ; before making the adjustment to
this parameter check the Home Offset of the unit
9.
In the menu Procedures select the “TEST SAMPLE INSERTION” end push exec.
.
•
If you have a problem can change the parameter of vescube.ini “INSERTION CLAMPV
CHAIN”,”INSETRION T VCHAIN OPEN CLAMP”,”INSERTION CLAMPV OVER CAP”,
“INSERTION T OVER CAP CLOSE CLAMP”, “INSERTION CLAMPV CAP DOWN”.
10.
From the “Procedures” section “TEST SAMPLE EXSTRACTION” end push exec.
•
If there is e problem whit the positioning (horizontal) of the clamp(ex when the clamp
go down it knock on the sensor) you must check the Home Offset of the unit.
•
If the clamp don’t correctly pick-up the tube from the chain(underneath the cap) you
can adjust the parameter “EXTRACTION CLAMPV CHAIN” for go further down it.
11.
Now Select the Procedure “TEST UNLOAD SAMPLE TO RACK” and press “Exec” .
•
If there is a problem you can adjust the parameter “CLAMP H RACK POS1” end
“UNLOAD T OPEN CLAMPV”.
12.
Exit from the Service menu by pressing “BACK”.
13.
Download the rack.
14.
Press the Start Button to execute the Reset procedure.
15.
Load a Classifier Rack, whit 5 samples, and when requested insert the ID code of the
rack, or read it with the handheld barcode reader.
16.
During the cycle check the correct working of the istrument:
a.
Identification/ Barcode reading
b.
Extraction/insertion of samples (clampV- clampH) (step 17)
c.
The correct work of mixer.
d.
Try to check the movement of the Reader Sensor (1 & 2) don’t must touch the
tube.
e.
Verify the correct positions of the Ejector.
17.
At the end of the exam press Stop, and thet press “Download Classif”.
18.
Check the printout with all the tubes in the classifier rack, comparing the barcodes
and the positions.
19.
Remove the two metallic keys (code 10345960) from the microswitches and place the
front panel in its correct position.
20.
Start the Analytical Startup procedure.