Installation Guide
27 of 33
Revision 2.00 of 04/11/2009
-
DELTA_LIMIT=11: Minimum derivative value used by Method 2
-
SH_THRESHOLD=100: Sample High detection threshold used by both
methods
-
START_POINT_M1=180: Right-most point analyzed by Method 1
-
END_POINT_M1=0: Left-most point analyzed by Method 1
-
START_POINT_M2=140: Right-most point analyzed by Method 2
-
END_POINT_M2=0: Left-most point analyzed by Method 2
-
DIFF_DIGITS=10: Value added to the LIMIT_VALUE to find the step in
Method 1
-
PERC_FILTER_MIN_ABS=20: Limiting filter applied in Method 1.
-
DIGITS_RATE=50: In method 1 as alternative condition to identify the
step
-
M1M2_MAX_DIFF_ESR=20: Maximum difference between the two ESR
values
-
PERC_CORR_ESR=60: Correction Value or ESR used by both methods
-
RETRY_VES=140: ESR final value above which the analysis is repeated
-
NUM_MAX_RETRY_VES=0: number of retry in case of error or values >
Retry_Ves
-
MAIN_METHOD=1: Definition of the Main Method
-
M1M2_MAX_DIFF_ERR=1: ERROR FLAG
-
SL_THRESHOLD=110: Sample Low Threshold.
-
START_POINT_M3=200: Right-most point analyzed by Method 3
-
END_POINT_M3=0: Left-most point analyzed by Method 3
T
he following section provides information about how to resolve some
problems.
We recommend, however, also consulting the manual to use the software
ReadViwer 1.
0..
When you change the parameters remember that END POINT M1 and
END POINT M2 ought to be same.
You can download the software at the following address :
Please note that access to the site is submitted to a password ; in case you do
not have your password please contact
After downloading unpack the file. Zip ; it is not necessary to create a specific
directory.
Evaluation :
Flag type
Max %
If < max %
If > max %
HIGH
2
ok
Check the tube
If the check results
Volume and
label