C U B E 3 0 T O U C H - U S E R M A N U A L
Rev 1.1 07/2018
13/56
-
Use appropriate personal protective equipment during operation.
In the event of biological material leakage during the operating cycle, clean the
outer surfaces of the instrument with 70% isopropyl alcohol and contact
Technical Services immediately.
External cleaning of the instrument
Periodic external cleaning is recommended.
1. With the instrument switched off, clean the instrument with a liquid
disinfectant used in the laboratory and leave it to dry. For the touch
screen use a dry microfibre cloth.
2. Repeat the operation with 70% isopropyl alcohol.
3. Leave the instrument turned off for at least 1 hour before starting a new
operating cycle or carrying out any other operation on the instrument.
Do not attempt to remove any screws or access the interior of the instrument.
For assistance with the interior of the instrument which is not readily accessible
to the operator, contact Technical Services.
DECONTAMINATION
PROCEDURE: