OPERATOR’S MANUAL
Instalab 700 NIR Product Analyzer
11001-1574-201107
SAMPLE PREPARATION 25
SEMI LIQUIDS AND SOLIDS
An open sample cup allows testing of semi liquids (pastes and slurries) and
solids (meats and cheeses). The open sample cup kit contains an open
sample cup and a sample cup holder.
To prepare Sample:
1. Place the sample cup holder onto a flat surface and put the sample cup
into the holder.
Figure 20
Sample Cup and Holder
2. Using a spatula or appropriate utensil, place the sample into the cup.
Figure 21
Preparing Sample Cup
3. The sample should be prepared such that the sample surface is flat
and flush with the rim of the cup.
Sample should be tightly packed to
eliminate any air pockets.
Sample Cup
Holder
Open
Sample
Cup
Summary of Contents for Instalab 700 NIR
Page 4: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 IV...
Page 6: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 2 SAFETY NOTICES...
Page 12: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 8 INTRODUCTION...
Page 38: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 34 NAVIGATION...
Page 70: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 66 ANALYZING PRODUCT...
Page 74: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 70 RESULTS...
Page 86: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 82 DIAGNOSTICS...
Page 96: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 92 MAINTENANCE...
Page 104: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 100 APPENDIX B...
Page 108: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 104 APPENDIX D...
Page 110: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 106 APPENDIX E...