OPERATOR’S MANUAL
ERROR 22
Probable Cause:
1. The measurement cell exceeded high limit.
Corrective Action:
1. Repeat test. If error recurs, contact DICKEY-john Service Department.
ERROR 23
Probable Cause:
1. The measurement cell weight is too low.
Corrective Action:
1. Increase the sample size and repeat the test.
ERROR 24
Probable Cause:
1. The measurement cell weight is too high.
Corrective Action:
1. Repeat the test. Be sure to follow display instructions, ensuring that
excess grain has been swept off the cell top.
ERROR 25
Probable Cause:
1. Error was encountered measuring weight.
Corrective Action:
1. Contact DICKEY-john Service Department.
ERROR 26
Probable Cause:
1. VF error is encountered.
Corrective Action:
1. Contact DICKEY-john Service Department.
ERROR 27 - 32
Probable Cause:
1. D1/D2 error is encountered.
Corrective Action:
1. Contact DICKEY-john Service Department.
32 TROUBLESHOOTING
GAC500XT
11001-1352-200510