OPERATOR’S MANUAL
ENTER
Enter button
Once empty cell measurements are complete, the display will prompt for
measurement (ENTER) or repeat of the empty cell (RIGHT ARROW). Press
the Enter button to measure D1 and D2, or the Right Arrow button to
repeat empty cell measurements. Press the Escape (ESC) button to return
to the Cell Test screen.
TEMPERATURE
The temperature test provides a measurement of cell temperature. The cell
may be empty or full for this measurement. Choose the TEMPERATURE
option from the Diagnostics screen. Press the Enter button. The
Temperature screen will be displayed (Figure
34).
Allow approximately 6
seconds for the unit to complete averaging measurements. The display will
update every 6 seconds with a new measurement. Press the Escape (ESC)
button to return to the Diagnostics screen.
Figure 34
Temperature Screen
Right Arrow button
ESC
Escape button
Temperature
F
81.29
Temp. Volts
1.20024
WEIGHT CHECK
The weight check feature may be used to measure the weight of the empty
cell or the grain currently held inside the cell. An empty cell test should be
conducted to verify the empty cell weighs 0.0 grams. Once this has been
completed, grain should be loaded into the hopper, then into the cell, and
the weight may be checked again. After the weight has been checked, the
excess grain in the grain drawer must be discarded, then the cell may be
emptied into the grain drawer. The weighed grain may then be taken to a
calibrated scale to verify the weight.
To access the weight check function, choose the WEIGHT CHECK option
from the Diagnostics screen. Press the Enter button. A message will be
displayed advising that the cell must be empty (Figure
35).
Figure 35
Weight Check Verification Screen
Weight Check
Make Sure Cell
is Empty And
Upright.
Then Press ENTER
26 DIAGNOSTICS
GAC500XT
11001-1352-200510
Summary of Contents for GAC500XT
Page 2: ...TABLE OF CONTENTS ii 11001 1352 200510 ...
Page 4: ...OPERATOR S MANUAL 2 SAFETY NOTICES GAC500XT 11001 1352 200510 ...
Page 10: ...OPERATOR S MANUAL 8 STARTUP GAC500XT 11001 1352 200510 ...
Page 18: ...OPERATOR S MANUAL 16 MEASURING MOISTURE GAC500XT 11001 1352 200510 ...
Page 20: ...OPERATOR S MANUAL 18 GRAIN SELECTION GAC500XT 11001 1352 200510 ...
Page 38: ...OPERATOR S MANUAL 36 TROUBLESHOOTING GAC500XT 11001 1352 200510 ...