Vacuum error
The measuring vacu-
um has been
dramatically
decreasing during
the measurement or
absolutely lost!
Leakage in the pneumatic system
(check the pump, fittings, tubings,
pressure sensor, puffer reservoir
and valves).
1106
Pneumatical error
occurred!
Device: %s
Error code: %s
Status: %x
Retry?
Fatal error in
mechanical
subsystem.
If frequently happens this error
check the proper operation of the
motor modules (home/end opto
switches, ribbon cables and
moving of motors).
1107
Check Cleaner container! Cleaner reagent
container became
empty, or the clean
fluid sensor is too
sensitive or there are
lot of bubbles in the
cleaner tubes!
Replace the Cleaner reagent
container with a filled one.
If this error still remains with a
filled cleaner reagent container,
calibrate sensors and check the
tubings and the fittings.
1113
Unrecoverable
pneumatical error
occurred!\n
Device: %s\n
Error code: %s\n
Status: %x\n
Please turn off the
instrument and turn it on
again!
Fatal error in
mechanical
subsystem. Possibly
mechanical jam.
Check the proper operation of the
motor modules (home/end opto
switches, ribbon cables and the
motor moving).
7.1.3. Measure errors
Code
Message
Reason(s)
Remedy
1200
Fatal “MeasInit” error.
Power off the system!
Fatal electronic error. Call service.
1201
Fatal HGB error.
Power off the system!
HGB channel did not
give a ready signal!
Call service.