5.2.10. Display and Keyboard Test
Display and keyboard test is provided to
check keypad and LCD panel.
Press a key to test and the LCD will show the
code of the pressed key, and will be invert the
colors at each touch of a key for testing that
every LCD dot is functioning.
5.2.11. Stress Mode
In
Stress mode,
the instrument performs measuring cycles without sample (blank
measurements) continuously. This can be used for burn-in tests, or to check pneumatic
system after changing any main fluidic parts.
You can have information about stability, cleanliness, HGB operation, and counting time
stability. Results of the last 10 PLT and HGB blank is displayed as well.
You can detect any kind of noise, or bubbles in the system if the PLT is not stable low, or
HGB has big variation. To exit from this mode
press the START button
(at the end of a
normal cycle) until the Stress operation is finished.