5.2.7. Service Testing Menu
Service Testing menu provides tools for checking hardware.
5.2.8. Valve Test Menu
In the menu on the left you can see the
valve numbers and a number under each
that represents the actual state of the
valve.
P means peristaltic pump
.
Selected valve number is shown in
inverse. 0 is
Off
state, 1 is
On
state.
Use cursor keys for selecting and setting
valve state, or press OK to toggle state.
5.2.9. Motor Test Menu
Pressing numeric key on keyboard the
regarding motors start operation repeatedly,
until pressing „1„ , or another key. All motors
stop at initial positions.