Omini o
perator’s
manual
version :19.04.17
5
3.5
D
ESCRIPTION OF THE CONTROL UNIT
PUMP
indicator turned on
Pump functioning: means the water purifier is either producing or
rinsing.
FLUSH
indicator turned on
The water purifier makes an automatic rinsing cycle of the reverse
osmosis membrane; a rinsing (of around 2 min) happens:
- following to the instrument starting,
- following a frequency of around 6 hours,
- following a pressure on the
MANUAL
button.
SOURCE
indicator turned on
Lack of flow or pressure on water supply.
FULL
indicator turned on
Full storage tank.
POWER
indicator turned on
water purifier under tension; this indicator is switched on
permanently, whatever the other indicators states are.
MANUAL
push button
Press this button makes starting a forced rinsing of the membrane
during 2 min.
water purifier producing
Rinsing cycle
Water supply default
Water storage full
Terms meaning :
PUMP : PUMP
FLUSH : RINSING
SOURCE : WATER SUPPLY
FULL : FULL TANK
POWER : UNDER TENSION
MANUAL : MANUAL RINSING