8
5.3
I
NSTALLATION OF THE WATER PURIFIER MEMBRANE
1.
Remove the membrane holder out of its two plastic stirrups . (Figure 1)
2.
Disconnect the water inlet pipe from the membrane holder by disconnecting the quick fit
coupling. (Figure 2)
3.
Unscrew (by turning to the right) the high streaked part ("big cap") of the membrane holder.
4.
Once the membrane door is opened. Insert the new membrane (Figure 3), peripheral seal at
the top, to the complete stop: the end of the collecting tube must be closed to the top of the
membrane holder. (Figure 4)
Figure 1
Figure 2
Figure 3
Figure 4
5.
Check that the O-ring is well-positioned at the bottom of the retaining wall of the membrane
holder (see the drawing).
6.
Unscrew (by turning to the right) the high streaked part ("big cap") of the membrane holder.
7.
Clip the membrane on its stirrups.