background image

 
 
 
 
 
 
 
 
 
 
 

Vet Photometer 

DP 700 

Operating Manual 

Version 5.9 

Edition 2020/2 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

                         

 

 
 

Summary of Contents for DP 700

Page 1: ...1 Vet Photometer DP 700 Operating Manual Version 5 9 Edition 2020 2 ...

Page 2: ...and specially designed for on site analysis With the new software version from V5 3 an automatic test of the instrument function has also been integrated With the Vet Photometer 4 parameters can be determined from serum plasma and blood NEFA free fatty acids calcium magnesium and lactate The kits and accessories required for the test are also available from Diaglobal GmbH All the best for your wor...

Page 3: ...oratory accessories 7 5 1 Durability of consumables 7 5 2 Reagents parameter list 8 5 3 Control materials 8 5 4 Laboratory aids 8 6 Internal quality control 9 7 Measuring process 9 7 1 Multipoint measurement and recognition of the end point 9 7 2 End point measurement taking standard 9 8 Measurement 10 8 1 Switching the instrument on 10 8 2 Self testing when switching on 10 8 3 Test selection 10 8...

Page 4: ...eter fulfils the basic requirements of Low Voltage Directive 2006 95 EG The conformity of the instrument with Directive 2006 95 EG is confirmed by the use of the CE marking The Vet Photometer was specially developed for analyses in veterinary medicine however it is universally applicable Manufacturer Diaglobal GmbH Innovationspark Wuhlheide Köpenicker Str 325 Haus 41 12555 Berlin Tel 49 0 30 6576 ...

Page 5: ...c waves Operation on a horizontal surface Please observe the following instructions for use Insert the rechargeable battery if the instrument is to be used network independent or connect the photometer to the power supply unit Press the key ON ENTER Fig 1 to activate the internal instrument check which is automatically carried out by the instrument Following this the instrument is immediately read...

Page 6: ...the normal battery using the push button contact and insert into the device Place the battery compartment cap back on top and fasten the knurled screws The rechargeable battery cannot be charged while it is installed A separate battery charger is required for this purpose Please note The Vet Photometer can be operated using a power supply and it is not necessary to remove the rechargeable battery ...

Page 7: ... cleaning the device with a soft cloth and the decontaminating solution it must be switched off and unplugged Make sure that no liquids get into the device There is no protection against penetrating liquids Code IP X0 The cuvette shaft must not be cleaned by the user of the device as this may damage the device Should it be necessary to clean it in particular in cases of leaking liquids or glass br...

Page 8: ... standard solution 2 x 10 mL 5 4 Disposables and accessories Art No Description Contents LH 004 Capillaries 10 µL with ringmark 250 LH 006 Cuvette rack 1 LH 007 Micropipettor pipetting aid 1 LH 009 Cellulose swabs 500 LH 010 Cellulose swab box 1 LH 011 Alcohol pads non sterile 100 LH 012 Powder free nitrile gloves size M 200 LH 050 Reaction tubes for samples 500 LH 053 Pipette tips 500 5000 µL cle...

Page 9: ...ped as soon as the end point is reached The time needed to reach the end point is temperature dependent It is normally 2 6 minutes for the lactate test If temperatures are close to freezing point measuring times of up to 20 minutes may result dependent on the parameters Both single measurements and series measurements can be selected If single measurements are made the samples are measured one aft...

Page 10: ... following display appears SERVICE In this case please switch off the instrument Please call the Service of Diaglobal GmbH Tel 49 0 30 6576 2597 or contact your speciality retailer 8 3 Test selection Press the ON ENTER key Select the desired method using from the menu using the right or left arrow key NEFA LAC CA MG ABS520 Press the right arrow key to mark the respective next test and the left arr...

Page 11: ...pensated Measurement range controls The measurement ranges of all measurement results indicated in the display become verified by an integrated measurement range control An error display appears when exceeding the measurement range All measurement ranges which are separately defined for each parameter are documented in the respective package inserts as well as in the operating manual chapter 9 Tec...

Page 12: ... width value 5 nm Extraneous light influence Negligible Interface V24 9600 8 n 2 Power supply 6 V 10 V DC Current consumption max 250 mA Warm up time 0 min Interference suppression According to DIN VDE 0871 and DIN VDE 0875 Electromagnetic Compatibility According to EN 61326 1 Inaccuracy 0 5 at A 1 000 Relative photometric short time standard deviation 0 1 Measuring ranges DP 700 Free fatty acids ...

Page 13: ...otometer meets the requirements of the IEC 61326 series of standards regarding electromagnetic radiation and interference immunity b Do not use this unit near to sources of intense electromagnetic radiation as these may disturb its correct functioning During the measurement a distance of at least 1 m should be kept between an operative turned on mobile telephone and this photometer Note on the uni...

Page 14: ...on Press ON ENTER Wait for device check and confirm with ON ENTER 2 Select test Press arrow key until required test appears 3 Confirm required test Press ON ENTER 4 Switch off Press both arrow keys at the same time Note If SERVICE appears after the device check the device has a fault In this case please contact our customer service at 49 0 30 6576 2597 ...

Page 15: ...N ENTER Wait for device check confirm with ON ENTER Select NEFA confirm with ON ENTER ON ENTER ON ENTER NEFA Insert blank 6 Insert cuvette 2 with standard into photometer Photometer shows Absorption of cuvette with standard Hint If necessary note this value for documentation purposes 7 Remove cuvette 2 with standard ON ENTER NEFA Sample 01 insert NEFA Insert standard ON ENTER NEFA Standard Abs 0 4...

Page 16: ...Absorption of cuvette with standard Hint If necessary note this value for documentation purposes 4 Blank adjustment Insert cuvette 1 blank into photometer blank is stored in memory Remove after signal tone ON ENTER ON ENTER CA Insert standard 6 Remove cuvette 2 with standard 7 Insert cuvette 3 sample 1 and thereafter cuvette 4 sample 2 into photometer Read results successively ON ENTER CA Sample 0...

Page 17: ...adjustment Insert cuvette with sample fig 3 into photometer blank is stored in memory Remove after signal tone 4 Switch on photometer with ON ENTER Wait for device check confirm with ON ENTER Select the required test confirm with ON ENTER 6 Replace screw cap against starter cap ON ENTER ON ENTER M1 LAC Sample 01 insert 9 Timing appears wait for result 7 Turn cuvette upside down several times 8 Fir...

Page 18: ...er blank adjustment of the last cuvette replace all screw caps against starter caps ON ENTER ON ENTER M1 LAC Sample 01 insert 8 Read the measured value of cuvette 1 remove cuvette Insert cuvette 2 read the measuring value remove cuvette etc 6 Turn cuvettes upside down several times at the same time 7 First press ON ENTER Thereafter insert cuvette 1 in photometer Timing appears wait for result ON E...

Reviews: