Step by step instructions
KRE 121
Single measurement
Additionally required: Dry block thermostat (30 minutes preheated)
Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · [email protected] · www.diaglobal.de 2021-01
1. Pipette 1 mL buffer solution
into each cuvette, then close
cuvettes
Note: The example shows 6
samples. They can only be
processed one after the other,
serial measurement is not possible
2. Incubate cuvettes in the
preheated dry block
thermostat for 7 minutes
3. Transfer 500 µL of the
sample (serum/plasma) into
the cuvette
Then insert the cuvette
immediately into the dry
block thermostat
5. After switching the
photometer on, wait for
device check and confirm
with ON/ENTER
Select CRE, confirm with
ON/ENTER
ON
ENTER
7. Wait for measured value
(2 minutes)
6. Insert cuvette into
photometer
Time is displayed
ON
ENTER
CRE
Sample 01 00:15
Measuring
CRE
Sample 01
insert
ON
ENTER
CRE
Sample 01
0.95 mg/dL
4. Incubate the cuvette with
the sample for exactly 1
minute
Already during the incubation
time, switch the photometer
on with ON/ENTER
8. Read the measured value,
remove the cuvette and
repeat the same process with
all other cuvettes, starting
from Fig. 3