Step by step instructions
HDL 321
Number of samples per series: Up to 20 samples at the same time
Additionally required: Mini centrifuge, Cholesterol CHO 142
Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · [email protected] · www.diaglobal.de 2021-01
1. HDL 321
Transfer 60 µL of the sample
with an end-to-end capillary
into each reaction tube "R"
and mix well
Leave it for 5 minutes
3. CHO 142
Insert capillary with sample
into cuvette
4. CHO 142
Eject sample several times with
micropipetter into cuvette
Screw cap on
Turn cuvette upside down several
times
6. CHO 142
Zero point adjustment: Insert
cuvette with sample (Fig. 4) into
photometer, zero point is stored
in memory
Remove cuvette after signal tone
5. CHO 142
Switch photometer on with ON/ENTER
key, wait for device check and
confirm with ON/ENTER
Select HDL/CHO, confirm with
ON/ENTER
Start with the measurement of CHO
7. CHO 142
Replace screw cap with
starter cap of CHO 142
ON
ENTER
ON
ENTER
M1 CHO
Sample 01
insert
9. CHO 142
First press ON/ENTER
Then insert cuvette into
photometer
Time is displayed, wait for
measured value
ON
ENTER
M2 CHO
Sample 01 00:30
Measuring
M1 CHO
Sample 02
insert or
ENTER
8. CHO 142
Turn cuvette upside down
several times
2. HDL 321
Insert reaction tube „R“ with
capillary into centrifuge
Centrifugate for 5 minutes
Continue with CHO 142