Step by step instructions
HDL 321
Number of samples per series: Up to 20 samples at the same time
Additionally required: Mini centrifuge, Cholesterol CHO 142
Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · [email protected] · www.diaglobal.de 2021-01
11. HDL 321
Pipette 500 µL supernatant from
the centrifuged reaction tube
"R" (Fig. 2) into the cuvette
HDL 321
12. HDL 321
Screw starter cap of HDL 321 on
15. HDL 321
Remove blank after signal tone
14. HDL 321
Zero point adjustment: Take an
unprocessed HDL 321 cuvette
(blank) from the package and
insert it into the photometer
Zero point is stored in memory
16. HDL 321
Insert cuvette with sample
(Fig. 13) in photometer
Read measured value
ON
ENTER
ON
ENTER
HDL
Insert
blank
HDL
Sample 01
insert
13. HDL 321
Turn cuvette upside down
several times
Wait for 5 minutes
10. CHO 142
The cholesterol value is stored
in memory
Now continue with the
measurement of HDL 321
Note: In order to carry out a
series measurement, all CHO 142
values must be measured first
All CHO 142 values will be stored
in memory, one after the other
Important: Pay attention to the
correct order and assignment of
the samples!