6.
Quality control according to the Guideline of the
German Medical Association
1)
The Vario Photometer II has been specially developed for near-patient immediate
diagnostics with unit-use reagents (German Medical Association, part B, chapter
2.1.5). According to the guideline of the German Medical Association, there is
therefore no obligation to participate in surveys (German Medical Association,
part B, chapter 2.2, paragraph (3) a). The user only has to carry out internal
quality checks.
Internal quality assurance is carried out in the form of a weekly accuracy check
(calibration) with subsequent documentation of the measured value. The
corresponding protocol forms are available from Diaglobal free of charge.
We recommend using the Diaglobal control solutions LAC QS and GLU QS to check
the accuracy of lactate and glucose determinations
.
We recommend using the blood control HEM QS and ERY QS with target values in
the normal concentration area for checking the accuracy of determinations of
haemoglobin, haematocrit and erythrocyte counts.
For all other parameters we recommend using the universal control sera from the
company, Roche, www.roche.de:
PreciControl ClinChem Multi 1 Order-No.: 05 947 626 190 (4 x 5 mL) for normal range
PreciControl ClinChem Multi 2 Order-No.: 05 947 774 190 (4 x 5 mL) for pathological range
In agreement with the requirements of the German Medical Association, a test of
the device function (see operating instructions, chapter 8.5) is integrated in the
Vario Photometer II, therefor a daily test by means of a standard manual test
(German Medical Association, part B, chapter 2.1.5, paragraph (2) is not
necessary.
The Vario Photometer II is suitable for the speedy detection of gestational
diabetes and fulfils the requirements of the Maternity Guidelines
2)
and the S3-
Directive
3)
. Glucose can be measured from whole blood as well as from venous
plasma. The displayed measured value is - according to the requirements - always
related to venous plasma.