Pag. 36 of 52
11.55 M55 - Analog output mode programming
In this menu it’s possible to set the analog output mode.
Default value: 0. 4-20mA
Available settings:
0. 4-20mA
1. 0-20mA
2. 0-20mA via RS232 (RS485)
3. 4-20mA vs.Sound
4. 20-4-20mA
5. 0-4-20mA
6. 20-0-20mA
7. 4-20mA vs.Vel.
8. 4-20mA vs.Energy
- Programming menu
11.56
M56 - 4mA (or 0mA) output programming
In this menu it’s possible to set analog output scale beginning.
The measure unit is in
M55
programming function, per es: with
M55
set to
“
0. 4-20mA
”, the measure unit is
m3/h
; with
M55
set to “
3. 4-20mA vs.Vel.
”
the measure unit is
m/s
.
11.57 M57 - 20mA output programming
In this menu it’s possible to set the analog output full scale.
The measure unit is in
M55
programming function, per es: with
M55
set to
“
0. 4-20mA
”, the measure unit is
m3/h
; with
M55
set to “
3. 4-20mA vs.Vel.
”
the measure unit is
m/s
.
11.58 M58 - Analog output simulation
In this menu it’s possible to force the analog output signal value to check
the drives connected to the 4÷20mA signal.
To start the simulation, press
and select with
or
the desired option
Available settings:
0
segnale in uscita 0mA
4
segnale in uscita 4mA
8
segnale in uscita 8mA
12 segnale in uscita 12mA
16 segnale in uscita 16mA
20 segnale in uscita 20mA
11.59 M59 - Analog output status displaying
In this menu it’s possible to display the analog output signal instantaneous
value.
NB
- It is not a measured value, but a value derived from a mathematical
calculation depending on the
M55
,
M56
and
M57
settings.
11.60 M60 - System date and time programming
In this menu it’s possible to impostare la data e l’ora del sistema
11.61 M61 - DIMPU info displaying
In this menu, the
DIMPU
unit details are available.