![DI MPU01 Manual Download Page 15](http://html1.mh-extra.com/html/di/mpu01/mpu01_manual_2492446015.webp)
Pag. 15 of 52
- Installation
7.6 Insertion transducer installation
Steps required for proper installation:
1 - with encased pipe, check that there is the minimum space required for the transducers installation (fig.23)
2 - with encased pipe, check that the free section length of the pipe is the minimum required (fig.24)
3 - Procuring the necessary equipment for drilling the pressure pipes
4 -
Set the pipe parameters: in the menu
M23
choose the option 17 or 21 (TC-1 or TLC-2 insertion transducer);
in the menu
M24
choose 1. (Z installation) and in the menu
M25
check the positioning distance
5 - Determine the best location for installation on pipe
6 - Installing the ball valve base
1. ultrasonic transducer signal emitter
2. ball valve base
3. ball valve
4. male thread for drill
5. sealing nut
6. head with terminals for electrical connection
7. connection cable
7.6.1 - Base valve Fixing
1. pipe
2. ball valve base
3. ball valve
4. drill bit Ø19mm
5. drill chuck
6. seal gland
7. drill rod
8. power drill
Fig. 24
Fig. 25
Fig. 26
Fig. 27
Fig. 28
Fig. 23