35
I F U
Immunity test
Basic EMC standard or
test method
IEC 60601-1-2 Edition 4 Test levels and compliance
Professional healthcare facility
environmen
t
Home healthcare environment
Surge
IEC 61000-4-5
± 0.5 kV, ± 1 kV
Voltage dips
IEC 61000-4-11
0% U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0% U
T
; 1 cycle
and
70% U
T
; 25/30 cycles.
Single phase: at 0°
Voltage Interruptions
IEC 61000-4-11
0% U
T
; 250 / 300 cycle
Conducted disturbanc-
es induced by RF fields
IEC 61000-4-6
6V in ISM bands between
0.15 MHz and 80 MHz
80% AM at 1 kHz
6V in ISM and amateur
bands between 0.15 MHz
and 80 MHz
80% AM at 1 kHz
Dwell time: 3 sec
Frequency step size: 1%
Radiated RF Electro-
magnetic fields
IEC 61000-4-3
3 V/m
80 MHz – 2.7 GHz
80% AM at 1 kHz
10 V/m
80 MHz – 2.7 GHz
80% AM at 1 kHz
Dwell time: 3 sec
Frequency step size: 1%
Electrical fast transients
/ burst
IEC 61000-4-4
± 2 kV
100 kHz repetition frequency
RATED Power frequen-
cy magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60Hz
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Contact: ± 8 kV
Air: ± 2 kV, ±4 kV, ± 8 kV, ± 15 kV
Number of discharges: 10 for each polarity